Shanghai Institute of Forensic Science, Shanghai Key Laboratory of Crime Scene Evidence, Shanghai 200083, China.
Forensic Sci Int. 2013 Dec 10;233(1-3):304-11. doi: 10.1016/j.forsciint.2013.10.004. Epub 2013 Oct 16.
A μElution solid-phase extraction (SPE) liquid chromatography-tandem mass spectrometry (LC-MS/MS) method for simultaneous determination of diazepam, nordiazepam, oxazepam, oxazepam glucuronide, temazepam and temazepam glucuronide in human whole blood is presented. 200 μL of whole blood samples were loaded onto a Waters Oasis HLB 96-well μElution SPE plate using 75 μL of methanol as the elution solvent, and the eluents were injected into an Eclipse XDB C18 column. No hydrolysis, solvent transfer, evaporation or reconstitution was involved in the sample preparation procedures. Tandem mass spectrometric detection with Turbo Ion Spray was conducted via multiple reaction monitoring (MRM) under positive ionization mode. The method was validated and proved to be accurate (accuracy within 93-108%), precise (intra-day RSD<9.9% and inter-day RSD<7.2%) and sensitive with limits of detection (LOD) in the range of 0.05-0.25 ng/mL for all the compounds. Extraction recoveries were in the range of 31-80% for all the analytes. This method demonstrated to be reproducible and reliable. The applicability of the method was demonstrated by analysis of several forensic cases involving diazepam and its metabolites.
本文建立了一种μElution 固相萃取(SPE)-液相色谱-串联质谱(LC-MS/MS)法,用于同时测定人全血中的地西泮、去甲西泮、奥沙西泮、奥沙西泮葡萄糖醛酸、替马西泮和替马西泮葡萄糖醛酸。将 200μL 全血样品加载到 Waters Oasis HLB 96 孔 μElution SPE 板上,用 75μL 甲醇作为洗脱溶剂,洗脱液注入 Eclipse XDB C18 柱。样品制备过程中无需水解、溶剂转移、蒸发或复溶。串联质谱检测采用 Turbo Ion Spray 在正离子模式下进行多反应监测(MRM)。该方法经过验证,结果准确(准确度在 93-108%之间)、精密度高(日内 RSD<9.9%,日间 RSD<7.2%),检测限(LOD)在 0.05-0.25ng/mL 范围内,所有化合物均具有较高的灵敏度。所有分析物的提取回收率在 31-80%之间。该方法重现性好,可靠性高。该方法已成功应用于几起涉及地西泮及其代谢物的法医案例分析。