Reverse shoulder glenoid loosening: an evaluation of the initial fixation associated with six different reverse shoulder designs.

作者信息

Stroud Nick, DiPaola Matthew J, Flurin Pierre-Henri, Roche Christopher P

出版信息

Bull Hosp Jt Dis (2013). 2013;71 Suppl 2:S12-7.

Abstract

INTRODUCTION

This study quantified glenoid fixation before and after cyclic loading of six reverse shoulder prosthesis designs when secured to low and high density bone substitute blocks.

METHODS

A displacement test quantified fixation of six reverse shoulder designs: 38 mm Equinoxe standard offset (EQ), 38 mm Equinoxe lateral offset (EQL), 36 mm Depuy Delta III (DRS), 36 mm Zimmer (ZRS), 32 mm DJO RSP (DJO), and a 36 mm Tornier BIO-RSA (BIO), secured to 0.24 and 0.48 g/cm 3 polyurethane blocks as a shear (357 N) and compressive (50 N) load were applied before and after cyclic loading. Displacement was measured with a dial indicator in the directions of the applied loads along the superior/inferior axis. A cyclic test rotated each glenosphere (N= 7) at 0.5 Hz for 10 k cycles as 750 N was constantly applied. A two-tailed Student's unpaired t-test compared mean displacements.

RESULTS

The average displacement of the EQ, EQL, ZRS, DJO, DRS, and BIO-RSA devices in the low density substrate was 182, 137, 431, 321, 190, and 256 microns, respectively. The average displacement of the EQ, EQL, ZRS, DRS, and BIO-RSA devices in the high density substrate was 102, 95, 244, 138, and 173 microns, respectively. Pre- and post-cyclic displacement was significantly less in the high density bone substitutes than in the low density bone substitutes for the majority of implant comparisons. During the cyclic test, six of seven ZRS devices failed at an average of 2,603 cycles, one of seven 32 mm DJO failed at 7,342 cycles, and four of seven BIO devices failed at an average of 2,926 cycles. All seven of the EQ, EQL, and DRS devices remained well fixed throughout cyclic loading.

DISCUSSION AND CONCLUSIONS

This study quantified glenoid fixation of six reverse shoulder designs; significant differences in fixation were observed between nearly every implant design tested.

摘要

引言

本研究对六种反向肩关节假体设计在固定于低密度和高密度骨替代块上时,循环加载前后的关节盂固定情况进行了量化。

方法

进行位移测试以量化六种反向肩关节设计的固定情况:38毫米Equinoxe标准偏移(EQ)、38毫米Equinoxe外侧偏移(EQL)、36毫米Depuy Delta III(DRS)、36毫米Zimmer(ZRS)、32毫米DJO RSP(DJO)和36毫米Tornier BIO-RSA(BIO),将其固定于0.24和0.48克/立方厘米的聚氨酯块上,在循环加载前后施加剪切力(357牛)和压缩力(50牛)。使用表盘指示器沿上下轴在施加载荷方向测量位移。进行循环测试,每个关节盂球体(N = 7)以0.5赫兹旋转10000次循环,同时持续施加750牛的力。采用双尾学生未配对t检验比较平均位移。

结果

在低密度基质中,EQ、EQL、ZRS、DJO、DRS和BIO-RSA装置的平均位移分别为182、137、431、321、190和256微米。在高密度基质中,EQ、EQL、ZRS、DRS和BIO-RSA装置的平均位移分别为102、95、244、138和173微米。在大多数植入物比较中,高密度骨替代物中循环前后的位移明显小于低密度骨替代物中的位移。在循环测试期间,七个ZRS装置中有六个平均在2603次循环时失效,七个32毫米DJO装置中有一个在7342次循环时失效,七个BIO装置中有四个平均在2926次循环时失效。所有七个EQ、EQL和DRS装置在整个循环加载过程中保持良好固定。

讨论与结论

本研究对六种反向肩关节设计的关节盂固定情况进行了量化;在所测试的几乎每种植入物设计之间均观察到固定存在显著差异。