Fuchs Katharina, Hezel Fabian, Klix Sabrina, Mekle Ralf, Wuerfel Jens, Niendorf Thoralf
Berlin Ultrahigh Field Facility (B.U.F.F.), Max-Delbrueck-Center for Molecular Medicine, Berlin, Germany.
Magn Reson Med. 2014 Dec;72(6):1590-8. doi: 10.1002/mrm.25066. Epub 2013 Dec 17.
This work proposes a dual contrast rapid acquisition with relaxation enhancement (RARE) variant (2in1-RARE), which provides simultaneous proton density (PD) and T2 * contrast in a single acquisition.
The underlying concept of 2in1-RARE is the strict separation of spin echoes and stimulated echoes. This approach offers independent weighting of spin echoes and stimulated echoes. 2in1-RARE was evaluated in phantoms including signal-to-noise ratio (SNR) and point spread function assessment. 2in1-RARE was benchmarked versus coherent RARE and a split-echo RARE variant. The applicability of 2in1-RARE for brain imaging was demonstrated in a small cohort of healthy subjects (n = 10) and, exemplary, a multiple sclerosis patient at 3 Tesla as a precursor to a broader clinical study.
2in1-RARE enables the simultaneous acquisition of dual contrast weighted images without any significant image degradation and without sacrificing SNR versus split-echo RARE. This translates into a factor of two speed gain over multi-contrast, sequential split-echo RARE. A 15% broadening of the point spread function was observed in 2in1-RARE. T1 relaxation effects during the mixing time can be neglected for brain tissue.
2in1-RARE offers simultaneous acquisition of images of anatomical (PD) and functional (T2 *) contrast. It presents an alternative to address scan time constraints frequently encountered during sequential acquisition of T2 * or PD-weighted RARE.
本研究提出了一种具有弛豫增强功能的双对比快速采集(RARE)变体(2in1-RARE),其可在单次采集中同时提供质子密度(PD)和T2*对比。
2in1-RARE的基本概念是自旋回波和受激回波的严格分离。这种方法可对自旋回波和受激回波进行独立加权。通过包括信噪比(SNR)和点扩散函数评估在内的体模对2in1-RARE进行了评估。将2in1-RARE与相干RARE和一种分割回波RARE变体进行了对比。在一小群健康受试者(n = 10)中证明了2in1-RARE对脑成像的适用性,并作为更广泛临床研究的前期工作,以3特斯拉场强对一名多发性硬化症患者进行了示例性研究。
2in1-RARE能够在不出现任何显著图像退化的情况下同时采集双对比加权图像,并且与分割回波RARE相比,不会牺牲SNR。这意味着与多对比、顺序分割回波RARE相比,速度提高了两倍。在2in1-RARE中观察到点扩散函数展宽了15%。对于脑组织,混合时间内的T1弛豫效应可忽略不计。
2in1-RARE可同时采集解剖学(PD)和功能性(T2*)对比的图像。它为解决在顺序采集T2*或PD加权RARE过程中经常遇到的扫描时间限制提供了一种替代方法。