From the Department of Biochemistry, University of Vermont, Colchester, Vermont.
J Trauma Acute Care Surg. 2014 Jan;76(1):107-13. doi: 10.1097/TA.0b013e3182aa80dc.
Viscoelastic measurements are frequently being used in clinical and research settings for a rapid assessment of the hemostatic processes of blood clot formation and degradation. These measurements are being performed on either of two instruments (TEG and ROTEM) using their proprietary reagents. Standardization between the instruments and the reagents has been lacking but is necessary to compare results across instruments. In this study, we perform a crossover analysis between the TEG and ROTEM instruments using proprietary reagents from each manufacturer.
We tested three sets of reagents as follows: (1) in-tem and ex-tem (Tem International GmbH); (2) kaolin and RapidTEG (Haemonetics); (3) a well-characterized control recombinant tissue factor-phospholipid reagent. Blood was drawn from six healthy donors, and each reagent was run concurrently in the TEG and ROTEM instruments. The volume of commercial reagent and calcium used was adjusted for crossover measurements to maintain the same concentration of each reagent in the blood. The outputs of clot time, rate of clot formation, and maximum firmness of the clot of the ROTEM and the TEG tracings were evaluated.
The in-tem and RapidTEG reagents showed no disparity between instruments for any parameter. Significant differences between the instruments were found in the α angle and maximum firmness of the clot for ex-tem and kaolin reagents as well as in the clot time and maximum firmness of the clot for the recombinant tissue factor-phospholipid reagent.
Although significant differences were observed for some parameters, the magnitudes were small compared with the differences between tests or the normal range variation in parameter values observed for these tests. These findings indicate that the instruments are more interchangeable than previously reported.
粘弹性测量常用于临床和研究环境中,以快速评估血液凝块形成和降解的止血过程。这些测量是在两种仪器(TEG 和 ROTEM)上使用其专用试剂进行的。仪器和试剂之间缺乏标准化,但为了比较仪器之间的结果,标准化是必要的。在这项研究中,我们使用每个制造商的专有试剂在 TEG 和 ROTEM 仪器之间进行交叉分析。
我们测试了三组试剂,如下所示:(1)in-tem 和 ex-tem(Tem International GmbH);(2)高岭土和 RapidTEG(Haemonetics);(3)一种经过充分表征的重组组织因子-磷脂试剂。从六名健康供体中抽取血液,并在 TEG 和 ROTEM 仪器上同时运行每种试剂。商业试剂和钙的体积进行了调整,以进行交叉测量,以保持血液中每种试剂的浓度相同。评估 ROTEM 和 TEG 轨迹的凝块时间、凝块形成速率和凝块最大硬度的输出。
in-tem 和 RapidTEG 试剂在任何参数方面仪器之间均无差异。ex-tem 和高岭土试剂的α角和凝块最大硬度以及重组组织因子-磷脂试剂的凝块时间和凝块最大硬度在仪器之间存在显著差异。
尽管一些参数存在显著差异,但与这些测试的测试间差异或参数值的正常范围变化相比,这些差异的幅度较小。这些发现表明仪器比以前报道的更可互换。