Starling 样流量控制左心室辅助装置的体外验证。

Starling-like flow control of a left ventricular assist device: in vitro validation.

机构信息

Division of Imaging Sciences and Biomedical Engineering, St. Thomas' Hospital, King's College London, London, UK.

出版信息

Artif Organs. 2014 Mar;38(3):E46-56. doi: 10.1111/aor.12221. Epub 2013 Dec 23.

Abstract

The application of rotary left ventricular (LV) assist devices (LVADs) is expanding from bridge to transplant, to destination and bridge to recovery therapy. Conventional constant speed LVAD controllers do not regulate flow according to preload, and can cause over/underpumping, leading to harmful ventricular suction or pulmonary edema, respectively. We implemented a novel adaptive controller which maintains a linear relationship between mean flow and flow pulsatility to imitate native Starling-like flow regulation which requires only the measurement of VAD flow. In vitro controller evaluation was conducted and the flow sensitivity was compared during simulations of postural change, pulmonary hypertension, and the transition from sleep to wake. The Starling-like controller's flow sensitivity to preload was measured as 0.39 L/min/mm Hg, 10 times greater than constant speed control (0.04 L/min/mm Hg). Constant speed control induced LV suction after sudden simulated pulmonary hypertension, whereas Starling-like control reduced mean flow from 4.14 to 3.58 L/min, maintaining safe support. From simulated sleep to wake, Starling-like control increased flow 2.93 to 4.11 L/min as a response to the increased residual LV pulsatility. The proposed controller has the potential to better match device outflow to patient demand in comparison with conventional constant speed control.

摘要

左心室(LV)辅助旋转设备(LVAD)的应用正在从桥接过渡到移植、目的地和桥接恢复治疗。传统的恒速 LVAD 控制器不能根据前负荷调节流量,可能导致过度/不足泵送,分别导致有害的心室抽吸或肺水肿。我们实施了一种新型的自适应控制器,该控制器保持平均流量和流量脉动之间的线性关系,以模拟原生的 Starling 样流量调节,只需要测量 VAD 流量。在体外进行了控制器评估,并在模拟体位变化、肺动脉高压和从睡眠到觉醒的过渡期间比较了流量灵敏度。Starling 样控制器对前负荷的流量灵敏度测量为 0.39 L/min/mm Hg,比恒速控制(0.04 L/min/mm Hg)高 10 倍。恒速控制在模拟突然发生的肺动脉高压后会导致 LV 抽吸,而 Starling 样控制会将平均流量从 4.14 降低至 3.58 L/min,从而保持安全支持。从模拟睡眠到觉醒,Starling 样控制会将流量从 2.93 增加到 4.11 L/min,以响应残留 LV 脉动性的增加。与传统的恒速控制相比,该提出的控制器具有更好地将设备输出与患者需求相匹配的潜力。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验