Jones D L, Sohla A, Bourland J D, Tacker W A, Kallok M J, Klein G J
Pacing Clin Electrophysiol. 1987 May;10(3 Pt 1):497-502. doi: 10.1111/j.1540-8159.1987.tb04512.x.
We compared single to sequential pulse shocks with different pulse separations on internal cardiac defibrillation by using a catheter and plaque electrodes in open-chest halothane-anesthetized pigs. Ten seconds after fibrillation onset, defibrillation was attempted using trapezoidal pulses of 65% tilt, approximately 5 ms duration and fixed outputs from 1.0 to 50 joules (J). With single pulses, minimum defibrillation energy for the catheter alone was 2.4 +/- 0.3 J/kg (mean +/- standard error) and 2.1 +/- 0.2 J/kg for the catheter tip to plaque configuration. With sequential pulse shocks, the first pulse delivered via the catheter and the second pulse from the catheter tip to the plaque electrode, the energy necessary for defibrillation was dependent on the separation time between the two pulses (2.0 +/- 0.2, 1.5 +/- 0.2, 0.9 +/- 0.1, 1.3 +/- 0.3, 0.6 +/- 0.2, and 1.2 +/- 0.2 J/kg at 100, 10, 1, 0.5, 0.2, and 0.1 ms, respectively). Further, at the 0.2 ms separation, 100% of the animals could be defibrillated with less than 2.0 J/kg (35 J total). We conclude that sequential pulse defibrillation provides a significant improvement over single pulse defibrillation. The optimum separation between the sequential pulses in this study was 0.2 ms.
我们在开胸氟烷麻醉的猪身上,使用导管和片状电极,比较了不同脉冲间隔下单次与连续脉冲电击对心脏内部除颤的效果。在心室颤动发作10秒后,尝试使用倾斜度为65%、持续时间约5毫秒、能量输出从1.0到50焦耳(J)固定的梯形脉冲进行除颤。对于单次脉冲,仅使用导管时的最小除颤能量为2.4±0.3焦耳/千克(平均值±标准误差),导管尖端至片状电极配置时为2.1±0.2焦耳/千克。对于连续脉冲电击,第一个脉冲通过导管输送,第二个脉冲从导管尖端传至片状电极,除颤所需能量取决于两个脉冲之间的间隔时间(在100、10、1、0.5、0.2和0.1毫秒时,分别为2.0±0.2、1.5±0.2、0.9±0.1、1.3±0.3、0.6±0.2和1.2±0.2焦耳/千克)。此外,在间隔0.2毫秒时,100%的动物可以用低于2.0焦耳/千克(总计35焦耳)的能量除颤。我们得出结论,连续脉冲除颤比单次脉冲除颤有显著改善。本研究中连续脉冲的最佳间隔为0.2毫秒。