Ostermeyer J, Horstkotte D, Bennett J, Huysmans H, Lindblom D, Olin C, Orinius E, Semb G
Thorac Cardiovasc Surg. 1987 Apr;35(2):71-7. doi: 10.1055/s-2007-1020200.
Between June 1980 and June 1983 4028 Björk-Shiley 70 degree convexo-concave prosthetic heart valves were distributed and implanted in Australia, Canada, Europe and South Africa. As of March 1986, a total of 52 outlet strut fractures (1.29%; 70% CL: 1.1%-1.5%) have been reported from 29 implant institutions in 12 countries. The majority (82.7%) occurred in Europe. Intervals between implantation and fracture were 13 days to 45.3 months (mean: 18.4 months; 70% CL: 16.6 months-20.1 months). The mortality rate after strut fracture was 78.7% (70% CL: 72.5%-84.9%). Upon stratification of the fracture by valve sizes and types it becomes evident that 75% (70% CL: 68.8%-81.2%) of all fractures are related to the sizes 29 mm to 33 mm (which virtually represent the same valve size) and predominantly to mitral valves (p less than 0.01). The large valves again have been stratified into two subsets, namely those fabricated from flanges originally machined as Björk-Shiley 60 degree convexo-concave valves (group I) and later produced valves machined initially to 70 degree specifications (group II). In group I the fracture rate was 5.2% (70% CL: 4.2%-6.2%) versus 1.6% (70% CL: 1.1%-2.1%) in group II (p less than 0.01), which identifies the group I 29 mm-33 mm Björk-Shiley 70 degree convexo-concave valves as the highest risk group for strut fracture. The rates are based upon all available information as of March 16, 1986.(ABSTRACT TRUNCATED AT 250 WORDS)
1980年6月至1983年6月期间,4028枚比约克-希利70度凸凹型人工心脏瓣膜在澳大利亚、加拿大、欧洲和南非分发并植入。截至1986年3月,12个国家的29个植入机构共报告了52例出口支柱骨折(1.29%;70%置信区间:1.1%-1.5%)。大多数(82.7%)发生在欧洲。植入与骨折之间的间隔为13天至45.3个月(平均:18.4个月;70%置信区间:16.6个月-20.1个月)。支柱骨折后的死亡率为78.7%(70%置信区间:72.5%-84.9%)。按瓣膜尺寸和类型对骨折进行分层后可以明显看出,所有骨折的75%(70%置信区间:68.8%-81.2%)与29毫米至33毫米尺寸的瓣膜有关(实际上代表相同的瓣膜尺寸),且主要与二尖瓣有关(p小于0.01)。大型瓣膜又被分为两个子集,即那些由最初加工为比约克-希利60度凸凹型瓣膜的法兰制成的瓣膜(第一组)和最初加工为70度规格的后期生产的瓣膜(第二组)。第一组的骨折率为5.2%(70%置信区间:4.2%-6.2%),而第二组为1.6%(70%置信区间:1.1%-2.1%)(p小于0.01),这表明第一组29毫米-33毫米的比约克-希利70度凸凹型瓣膜是支柱骨折的最高风险组。这些发生率是基于截至1986年3月16日的所有可用信息。(摘要截取自250字)