Department of Oral and Maxillofacial Radiology, Peking University School and Hospital of Stomatology, Beijing, China.
Dentomaxillofac Radiol. 2014;43(3):20130414. doi: 10.1259/dmfr.20130414. Epub 2014 Jan 22.
To assess the impact of spatial resolution and cone beam CT (CBCT) unit on CBCT images for the detection accuracy of condylar defects.
42 temporomandibular joints were scanned, respectively, with the CBCT units ProMax® 3D (Planmeca Oy, Helsinki, Finland) and DCT PRO (Vatech, Co., Ltd., Yongin-Si, Republic of Korea) at normal and high resolutions. Seven dentists evaluated all the test images with respect to the presence or the absence of condylar defects. Receiver operating characteristic curve analysis was employed to define the detection accuracy. Two-way analysis of variance was used to analyse the values under the receiver operating characteristic curves for the differences among imaging groups and observers. Intraobserver variation was analysed using the Wilcoxon test.
Macroscopic anatomy examination revealed that, of the 42 temporomandibular joint condylar surfaces, 18 were normal and 24 had defects on the surface of condyles. No significant differences were found between the images scanned with normal and high resolutions for both CBCT units ProMax 3D (p = 0.119) and DCT PRO (p = 0.740). Significant differences exist between image groups of DCT PRO and ProMax 3D (p < 0.05). Neither the inter- nor the intraobserver variability were significant.
The spatial resolution per se did not have an impact on the detection accuracy of condylar defects. The detection accuracy of condylar defects highly depends on the CBCT unit used for examination.
评估空间分辨率和锥形束 CT(CBCT)对髁突缺损检测准确性的 CBCT 图像的影响。
分别使用 ProMax® 3D(Planmeca Oy,赫尔辛基,芬兰)和 DCT PRO(Vatech,Co.,Ltd.,Yongin-Si,韩国)CBCT 设备以正常和高分辨率扫描 42 个颞下颌关节。7 名牙医评估了所有测试图像中髁突是否存在缺陷。采用受试者工作特征曲线分析来定义检测准确性。采用双向方差分析来分析不同成像组和观察者之间的受试者工作特征曲线下值。采用 Wilcoxon 检验分析观察者内变异性。
宏观解剖学检查显示,在 42 个颞下颌关节髁突表面中,18 个正常,24 个表面有缺陷。ProMax 3D(p=0.119)和 DCT PRO(p=0.740)两种 CBCT 设备的正常和高分辨率扫描图像之间没有发现差异。DCT PRO 和 ProMax 3D 的图像组之间存在显著差异(p<0.05)。观察者内和观察者间的变异性均不显著。
空间分辨率本身不会影响髁突缺损的检测准确性。髁突缺陷的检测准确性高度依赖于用于检查的 CBCT 设备。