Van Linden Arnaud, Kempfert Jörg, Blumenstein Johannes, Möllmann Helge, Kim Won-Keun, Alkaya Serap, Hamm Christian, Walther Thomas
Department of Cardiac Surgery, Kerckhoff-Klinik, Bad Nauheim, Germany
Department of Cardiac Surgery, Kerckhoff-Klinik, Bad Nauheim, Germany.
Eur J Cardiothorac Surg. 2014 Aug;46(2):207-12; discussion 212. doi: 10.1093/ejcts/ezt600. Epub 2014 Jan 14.
Precise measurements of aortic annulus dimensions are crucial for prosthesis size selection in patients undergoing transcatheter aortic valve implantation (TAVI). The so-called effective diameter (derived from area) measured in multislice computed tomography (MSCT) images has evolved to be the most precise measurement tool. Usually, the operator must manually adjust the aortic annular plane. Syngo Aortic ValveGuide (Siemens Healthcare) is a new software tool that allows for automatic aortic root reconstruction and annular plane detection. The aim of this study was to compare the effective diameter measured in automatically detected and manually adjusted annular plane.
Seventy-three raw image datasets of preoperative TAVI MSCT scans were analysed using our institutional gold standard (3Mensio Valves™) with manual annular plane adjustment and using Aortic ValveGuide with automatic annular plane detection. The aortic annular circumference was manually marked for both software tools, and the effective diameter was calculated using the formula: effective diameter = 2 × √(circumferential area/π).
Automatic annulus plane detection using Syngo Aortic ValveGuide worked well in all MSCT scans. Minor manual adjustment of the detected plane was necessary in only 3 patients. The mean effective aortic annulus diameter was 23.1 ± 2.4 mm for 3Mensio and 23.3 ± 2.4 mm for Syngo Aortic ValveGuide. Bland-Altman analysis of both imaging software tools showed good agreement (mean difference of 0.16 mm and limits of agreement of -0.48 to 0.80 mm).
Effective aortic annulus diameter measured with Syngo Aortic ValveGuide, as a new imaging software that allows for automatic aortic annular plane detection, shows good agreement to gold standard measurements. Automatic annulus plane detection might reduce the effort for MSCT analysis and may lead to more reproducible aortic annulus measurements. Aortic ValveGuide is part of the DynaCT and in future aortic annulus dimension measurements may be feasible during intraoperatively acquired DynaCT.
精确测量主动脉瓣环尺寸对于接受经导管主动脉瓣植入术(TAVI)的患者选择合适的假体尺寸至关重要。多层螺旋计算机断层扫描(MSCT)图像中测量的所谓有效直径(由面积推导得出)已逐渐成为最精确的测量工具。通常,操作人员必须手动调整主动脉瓣环平面。Syngo主动脉瓣指南(西门子医疗)是一种新的软件工具,可实现主动脉根部的自动重建和瓣环平面检测。本研究的目的是比较在自动检测和手动调整的瓣环平面中测量的有效直径。
使用我们机构的金标准(3Mensio瓣膜™)并手动调整瓣环平面,以及使用具有自动瓣环平面检测功能的主动脉瓣指南,对73例术前TAVI MSCT扫描的原始图像数据集进行分析。两种软件工具均手动标记主动脉瓣环周长,并使用公式:有效直径 = 2×√(圆周面积/π)计算有效直径。
使用Syngo主动脉瓣指南进行自动瓣环平面检测在所有MSCT扫描中均效果良好。仅3例患者需要对检测到的平面进行轻微手动调整。3Mensio测量的平均有效主动脉瓣环直径为23.1±2.4mm,Syngo主动脉瓣指南测量的为23.3±2.4mm。两种成像软件工具的Bland-Altman分析显示一致性良好(平均差异为0.16mm,一致性界限为-0.48至0.80mm)。
作为一种能够自动检测主动脉瓣环平面的新型成像软件,使用Syngo主动脉瓣指南测量的有效主动脉瓣环直径与金标准测量结果显示出良好的一致性。自动瓣环平面检测可能会减少MSCT分析的工作量,并可能导致更可重复的主动脉瓣环测量。主动脉瓣指南是DynaCT的一部分,未来在术中获取的DynaCT期间进行主动脉瓣环尺寸测量可能是可行的。