Department of Chemistry and Biochemistry, University of Notre Dame, Notre Dame, IN 46556, USA.
Department of Chemistry and Biochemistry, University of Notre Dame, Notre Dame, IN 46556, USA.
Anal Chim Acta. 2014 Jan 31;810:94-8. doi: 10.1016/j.aca.2013.11.057. Epub 2013 Dec 7.
A PrinCE autosampler was coupled to a Q-Exactive mass spectrometer by an electrokinetically pumped sheath-flow nanospray interface to perform automated capillary zone electrophoresis-electrospray ionization-tandem mass spectrometry (CZE-ESI-MS/MS). 20ng aliquots of an Escherichia coli digest were injected to evaluate the system. Eight sequential injections over an 8-h period identified 1115±70 (relative standard deviation, RSD=6%) peptides and 270±8 (RSD=3%) proteins per run. The average RSDs of migration time, peak intensity, and peak area were 3%, 24% and 19%, respectively, for 340 peptides with high intensity. This is the first report of an automated CZE-ESI-MS/MS system using the electrokinetically pumped sheath-flow nanospray interface. The results demonstrate that this system is capable of reproducibly identifying over 1000 peptides from an E. coli tryptic digest in a 1-h analysis time.
采用电动泵鞘流纳喷雾接口将 PrinCE 自动进样器与 Q-Exactive 质谱仪相连接,以进行自动化毛细管区带电泳-电喷雾串联质谱(CZE-ESI-MS/MS)分析。进样 20ng 的大肠杆菌消化物以评估该系统。在 8 小时的时间内进行 8 次连续进样,每个运行可鉴定出 1115±70(相对标准偏差,RSD=6%)个肽段和 270±8(RSD=3%)个蛋白。对于具有高强度的 340 个肽段,迁移时间、峰强度和峰面积的平均 RSD 分别为 3%、24%和 19%。这是首次报道采用电动泵鞘流纳喷雾接口的自动化 CZE-ESI-MS/MS 系统。结果表明,该系统能够在 1 小时的分析时间内,可重复性地从大肠杆菌胰蛋白酶消化物中鉴定出 1000 多个肽段。