Dall-Miles 钢缆固定系统治疗转子间骨折早期灾难性失败。

Early catastrophic failure of trochanteric fixation with the Dall-Miles Cable Grip System.

机构信息

Department of Orthopaedic Surgery, The Warren Alpert Medical School of Brown University, Providence, Rhode Island.

出版信息

J Arthroplasty. 2014 Jun;29(6):1289-91. doi: 10.1016/j.arth.2014.01.001. Epub 2014 Jan 10.

Abstract

Multi-filament orthopaedic cerclage cables have been used in clinical practice for several decades, and are commonly utilized to provide greater trochanteric fixation following an osteotomy or fracture. We present the first known report of patients who experienced early catastrophic failure following use of the Dall-Miles Cable Grip System (DMCGS). A root cause analysis determined that the jaws of one crimper had an increased distance at closure, resulting in inadequate ultimate fixation strength. It was discovered that operating room staff had not been trained to perform regular calibration checks and the required calibration tool was not included in any of the institution's DMCGS sets. Surgeons should be aware that these surgical instruments require regular maintenance and should be gauge tested prior to every use.

摘要

多股骨科环形结扎线在临床实践中已经使用了几十年,通常用于在截骨或骨折后提供更大转子固定。我们报告了首例使用 Dall-Miles Cable Grip System(DMCGS)后早期灾难性失效的患者。根本原因分析确定,一个压接钳的钳口在闭合时的距离增加,导致最终固定强度不足。发现手术室工作人员没有接受过定期校准检查的培训,而且机构的任何一套 DMCGS 中都没有包含所需的校准工具。外科医生应该意识到这些手术器械需要定期维护,并在每次使用前进行测量测试。

相似文献

1
Early catastrophic failure of trochanteric fixation with the Dall-Miles Cable Grip System.
J Arthroplasty. 2014 Jun;29(6):1289-91. doi: 10.1016/j.arth.2014.01.001. Epub 2014 Jan 10.
2
Early experience with a novel nonmetallic cable in reconstructive hip surgery.
Clin Orthop Relat Res. 2010 Sep;468(9):2382-6. doi: 10.1007/s11999-010-1284-x.
5
Trochanteric fixation by cable grip in hip replacement.
J Bone Joint Surg Br. 1991 Jul;73(4):580-1. doi: 10.1302/0301-620X.73B4.2071639.
8
Cerclage Breakage in Trochanteric Revision - High Risk for Hip Endoprosthesis?
Acta Chir Orthop Traumatol Cech. 2017;84(2):138-140.
9
Fixation systems of greater trochanteric osteotomies: biomechanical and clinical outcomes.
J Am Acad Orthop Surg. 2007 Oct;15(10):614-24. doi: 10.5435/00124635-200710000-00006.
10
Transfemoral implantation of the Wagner SL stem. The abolition of subsidence and enhancement of osteotomy union rate using Dall-Miles cables.
Arch Orthop Trauma Surg. 2002 Dec;122(9-10):557-60. doi: 10.1007/s00402-002-0424-5. Epub 2002 Sep 13.

引用本文的文献

1
Application of a Fork Plate for Greater Trochanter Osteosynthesis in Total Hip Arthroplasty.
Sovrem Tekhnologii Med. 2020;12(2):80-84. doi: 10.17691/stm2020.12.2.10.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验