The use and interaction of permanent pacemakers and the automatic implantable cardioverter defibrillator.

作者信息

Cohen A I, Wish M H, Fletcher R D, Miller F C, McCormick D, Shuck J, Shapira N, Delnegro A A

机构信息

Cardiology service, Veteran's Administration Medical Center, Washington, DC.

出版信息

Pacing Clin Electrophysiol. 1988 Jun;11(6 Pt 1):704-11. doi: 10.1111/j.1540-8159.1988.tb06020.x.

Abstract

The adverse interactions of permanent pacemakers and automatic implantable cardioverter defibrillators (AICD) were studied in nine patients in whom both devices were implanted. Both unipolar and bipolar pacemakers were evaluated. The permanent pacemakers were also used to do noninvasive electrophysiological studies and to induce ventricular fibrillation. Undersensing of ventricular fibrillation by the permanent pacemakers caused inappropriate pacemaker stimuli, which caused undersensing of ventricular fibrillation by the AICD in three of four patients with unipolar pacemakers. After an AICD discharge, pacemaker noncapture was seen in eight of 22 episodes for an average 4.9 seconds and inability to sense was seen in 11 of 20 episodes for an average 9.0 seconds. Counting of pacemaker stimuli and QRS by the AICD caused inappropriate discharges. Noninvasive electrophysiological testing by the pacemakers correlated with invasive testing. Furthermore, induction of ventricular fibrillation was successful in four of five patients attempted, though requiring long bursts at high outputs at the shortest cycle lengths obtainable by these pacemakers. Operation of the AICD and permanent pacemakers must be clearly understood to avoid adverse interactions of these devices.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验