Santos Joana, Batista Maria do Carmo, Foley Shane, Paulo Graciano, McEntee Mark F, Rainford Louise
Instituto Politécnico de Coimbra, ESTESC-Coimbra Health School, Radiologia, Rua 5 de Outubro, S. Martinho do Bispo, 3046-854 Coimbra, Portugal
Departamento de Física Médica, Dr. Campos Costa, Consultório de Tomografia Computorizada S.A, Porto, Portugal.
Radiat Prot Dosimetry. 2014 Dec;162(4):586-96. doi: 10.1093/rpd/ncu018. Epub 2014 Feb 23.
The purpose of the study is to perform phantom-based optimisation of paediatric computed tomography (CT) protocols and quantify the impact upon radiation dose and image noise levels. The study involved three Portuguese paediatric centres. Currently employed scanning protocols for head and chest examinations and combinations of exposure parameters were applied to a Catphan(®)600 phantom to review the CT dose impact. Contrast-noise ratio (CNR) was quantified using Radia Diagnostic(®) tool. Imaging parameters, returning similar CNRs (<1) and dose savings were applied to three paediatric anthropomorphic phantoms. OsiriX software based on standard deviation pixel values facilitated image noise analysis. Currently employed protocols and age categorisation varied between centres. Manipulation of exposure parameters facilitated mean dose reductions of 33 and 28 % for paediatric head and chest CT examinations, respectively. The majority of the optimised CT examinations resulted in image noise similar to currently employed protocols. Dose reductions of up to 33 % were achieved with image quality maintained.
本研究的目的是基于体模对儿科计算机断层扫描(CT)协议进行优化,并量化其对辐射剂量和图像噪声水平的影响。该研究涉及三个葡萄牙儿科中心。将目前用于头部和胸部检查的扫描协议以及曝光参数组合应用于Catphan(®)600体模,以评估CT剂量影响。使用Radia Diagnostic(®)工具对对比度噪声比(CNR)进行量化。将返回相似CNR(<1)且节省剂量的成像参数应用于三个儿科仿真人体模型。基于标准差像素值的OsiriX软件有助于图像噪声分析。各中心目前采用的协议和年龄分类各不相同。通过调整曝光参数,儿科头部和胸部CT检查的平均剂量分别降低了33%和28%。大多数优化后的CT检查产生的图像噪声与目前使用的协议相似。在保持图像质量的同时,剂量降低了33%。