Santos Joana, Foley Shane, Paulo Graciano, McEntee Mark F, Rainford Louise
Instituto Politécnico de Coimbra, ESTESC, DMIR, Rua 5 de Outubro, S. Martinho do Bispo, 3046-854, Coimbra, Portugal.
School of Medicine & Medical Science, Health Science Centre, University College Dublin, Belfield, Dublin 4, Ireland.
Pediatr Radiol. 2015 Nov;45(12):1814-22. doi: 10.1007/s00247-015-3398-5. Epub 2015 Aug 5.
The volume of CT examinations has increased with resultant increases in collective dose values over the last decade.
To analyze the impact of the tube current and voltage modulation for dose values and image quality of pediatric head CT examinations.
Head CT examinations were performed on anthropomorphic phantoms and four pediatric age categories before and after the introduction of dedicated pediatric curves for tube voltage and current modulation. Local diagnostic reference levels were calculated. Visual grading characteristic image quality evaluation was performed by four pediatric neuroradiologists and image noise comparisons were performed.
Pediatric-specific modulation curves demonstrated a 49% decrease in mean radiation dose for phantom examinations. The local diagnostic reference levels (CTDIvol) for clinical examinations decreased by 52%, 41%, 46% and 40% for newborn, 5-, 10- and 15-year-old patients, respectively. Visual grading characteristic image quality was maintained for the majority of age categorizations (area under the curve = 0.5) and image noise measurements did not change (P = 0.693).
Pediatric-specific dose modulation curves resulted in an overall mean dose reduction of 45% with no significant differences in subjective or objective image quality findings.
在过去十年中,CT检查的数量有所增加,导致集体剂量值上升。
分析管电流和电压调制对小儿头部CT检查的剂量值和图像质量的影响。
在引入专门的小儿管电压和电流调制曲线之前和之后,对仿真人体模型和四个小儿年龄组进行头部CT检查。计算局部诊断参考水平。由四位小儿神经放射科医生进行视觉分级特征图像质量评估,并进行图像噪声比较。
小儿专用调制曲线显示,仿真人体模型检查的平均辐射剂量降低了49%。临床检查的局部诊断参考水平(CTDIvol),新生儿、5岁、10岁和15岁患者分别降低了52%、41%、46%和40%。大多数年龄组的视觉分级特征图像质量得以维持(曲线下面积=0.5),图像噪声测量无变化(P=0.693)。
小儿专用剂量调制曲线使总体平均剂量降低了45%,主观或客观图像质量结果无显著差异。