Suppr超能文献

[用于水吸收剂量(N(D,w))刻度的治疗级剂量计校准的⁶⁰Coγ射线标准场的研制]

[Development of the 60Co gamma-ray standard field for therapy-level dosimeter calibration in terms of absorbed dose to water (N(D,w))].

作者信息

Fukumura Akifumi, Mizuno Hideyuki, Fukahori Mai, Sakata Suoh

机构信息

National Institute of Radiological Sciences, Chiba, Japan.

出版信息

Igaku Butsuri. 2012;32(4):182-8.

Abstract

A primary standard for the absorbed dose rate to water in a 60Co gamma-ray field was established at National Metrology Institute of Japan (NMIJ) in fiscal year 2011. Then, a 60Co gamma-ray standard field for therapy-level dosimeter calibration in terms of absorbed dose to water was developed at National Institute of Radiological Sciences (NIRS) as a secondary standard dosimetry laboratory (SSDL). The results of an IAEA/WHO TLD SSDL audit demonstrated that there was good agreement between NIRS stated absorbed dose to water and IAEA measurements. The IAEA guide based on the ISO standard was used to estimate the relative expanded uncertainty of the calibration factor for a therapy-level Farmer type ionization chamber in terms of absorbed dose to water (N(D,w)) with the new field. The uncertainty of N(D,w) was estimated to be 1.1% (k = 2), which corresponds to approximately one third of the value determined in the existing air kerma field. The dissemination of traceability of the calibration factor determined in the new field is expected to diminish the uncertainty of dose delivered to patients significantly.

摘要

2011财年,日本国家计量院(NMIJ)建立了60Coγ射线场中水中吸收剂量率的一级标准。随后,作为二级标准剂量学实验室(SSDL),日本国立放射科学研究所(NIRS)开发了用于治疗级剂量计水吸收剂量校准的60Coγ射线标准场。国际原子能机构/世界卫生组织热释光剂量计SSDL审核结果表明,NIRS公布的水吸收剂量与国际原子能机构的测量结果吻合良好。基于ISO标准的国际原子能机构指南用于估算新场中治疗级 Farmer 型电离室水吸收剂量校准因子(N(D,w))的相对扩展不确定度。N(D,w)的不确定度估计为1.1%(k = 2),约为现有空气比释动能场中确定值的三分之一。预计新场中确定的校准因子的溯源性传播将显著降低输送给患者的剂量的不确定度。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验