Morishita Y, Kato M, Takata N, Kurosawa T, Tanaka T, Saito N
National Metrology Institute of Japan, AIST, Tsukuba, Ibaraki 305-8568, Japan.
Radiat Prot Dosimetry. 2013;154(3):331-9. doi: 10.1093/rpd/ncs235. Epub 2012 Sep 5.
A primary standard for the absorbed dose rate to water in a ⁶⁰Co radiation field has been newly established at the National Metrology Institute of Japan. This primary standard combines the calorimetric measurements using a graphite calorimeter with the ionometric measurements using a thick-walled graphite cavity ionisation chamber. The calorimeter is operated in the constant temperature mode using AC Wheatstone bridges. The absorbed dose rate to water was determined to be 12 mGy s⁻¹ at a point of 1 m from the radiation source and at a water depth of 5 g cm⁻². The uncertainty on the calibration coefficient in terms of the absorbed dose to water of an ionisation chamber using this standard was estimated to be 0.39 % (k=1).
日本国家计量院新建立了钴-60辐射场中水吸收剂量率的基准标准。该基准标准将使用石墨量热计的量热测量与使用厚壁石墨腔电离室的电量测量相结合。量热计使用交流惠斯通电桥在恒温模式下运行。在距辐射源1米处且水深为5克/平方厘米的点处,水吸收剂量率被确定为12毫戈瑞/秒。使用该标准对电离室水吸收剂量校准系数的不确定度估计为0.39%(k=1)。