Müller-Lutz Anja, Khalil Nadia, Schmitt Benjamin, Jellus Vladimir, Pentang Gael, Oeltzschner Georg, Antoch Gerald, Lanzman Rotem S, Wittsack Hans-Jörg
Department of Diagnostic and Interventional Radiology, Medical Faculty, University Dusseldorf, Moorenstrasse 5, 40225, Düsseldorf, Germany,
MAGMA. 2014 Dec;27(6):477-85. doi: 10.1007/s10334-014-0433-8. Epub 2014 Feb 26.
The objective of this study was to show the feasibility to perform Iopamidol-based pH imaging via clinical 3T magnetic resonance imaging (MRI) using chemical exchange saturation transfer (CEST) imaging with pulse train presaturation.
The pulse train presaturation scheme of a CEST sequence was investigated for Iopamidol-based pH measurements using a 3T magnetic resonance (MR) scanner. The CEST sequence was applied to eight tubes filled with 100-mM Iopamidol solutions with pH values ranging from 5.6 to 7.0. Calibration curves for pH quantification were determined. The dependence of pH values on the concentration of Iopamidol was investigated. An in vivo measurement was performed in one patient who had undergone a previous contrast-enhanced computed tomography (CT) scan with Iopamidol. The pH values of urine measured with CEST MRI and with a pH meter were compared.
In the measured pH range, pH imaging using CEST imaging with pulse train presaturation was possible. Dependence between the pH value and the concentration of Iopamidol was not observed. In the in vivo investigation, the pH values in the human bladder measured by the Iopamidol CEST sequence and in urine were consistent.
Our study shows the feasibility of using CEST imaging with Iopamidol for quantitative pH mapping in vitro and in vivo on a 3T MR scanner.
本研究的目的是展示通过临床3T磁共振成像(MRI),利用带有脉冲序列预饱和的化学交换饱和转移(CEST)成像来进行基于碘帕醇的pH成像的可行性。
使用3T磁共振(MR)扫描仪,研究了CEST序列的脉冲序列预饱和方案用于基于碘帕醇的pH测量。将CEST序列应用于8支装有pH值范围为5.6至7.0的100 mM碘帕醇溶液的试管。确定了用于pH定量的校准曲线。研究了pH值对碘帕醇浓度的依赖性。对一名曾接受过碘帕醇增强计算机断层扫描(CT)的患者进行了体内测量。比较了用CEST MRI和pH计测量的尿液pH值。
在测量的pH范围内,使用带有脉冲序列预饱和的CEST成像进行pH成像是可行的。未观察到pH值与碘帕醇浓度之间的依赖性。在体内研究中,通过碘帕醇CEST序列测量的人膀胱中的pH值与尿液中的pH值一致。
我们的研究表明,在3T MR扫描仪上,使用碘帕醇的CEST成像在体外和体内进行定量pH映射是可行的。