Wagatsuma Kei, Miwa Kenta, Akimoto Kenta, Tsushima Hiroyuki, Miyaji Noriaki, Umeda Takuro, Murata Taisuke, Takiguchi Tomohiro, Koizumi Mitsuru
Department of Nuclear Medicine, Cancer Institute Hospital of Japanese Foundation for Cancer Research.
Nihon Hoshasen Gijutsu Gakkai Zasshi. 2014 Feb;70(2):148-53. doi: 10.6009/jjrt.2014_jsrt_70.2.148.
Quality control (QC) detects changes in the performance of gamma cameras that could adversely affect interpretations of clinical studies. We used plate and sheet (57)Co flood sources to measure extrinsic uniformity during daily QC. Each source, when placed on the top of a collimated detector, allowed the acquisition of uniform images from both detectors, thus reducing the amount of time needed to perform daily QC. No serious problems with the gamma camera system were revealed by visual checks, and changes in detector sensitivity were rapidly determined by observing daily variations in the measured values of extrinsic uniformity. Furthermore, (57)Co flood sources confer advantages in that they shorten the time required for preparation of flood sources and reduce the consequent exposure of medical staff to radiation.
质量控制(QC)可检测γ相机性能的变化,这些变化可能会对临床研究的解读产生不利影响。我们使用平板和薄片(57)钴泛源在每日质量控制期间测量外部均匀性。每个源放置在准直探测器顶部时,可从两个探测器获取均匀图像,从而减少了进行每日质量控制所需的时间。通过目视检查未发现γ相机系统存在严重问题,并且通过观察外部均匀性测量值的每日变化,可快速确定探测器灵敏度的变化。此外,(57)钴泛源具有缩短泛源准备所需时间并减少医护人员因此受到辐射暴露的优点。