Sheykhi-Dolagh Roghaye, Saeedi Hassan, Farahmand Behshid, Kamyab Mojtaba, Kamali Mohammad, Gholizadeh Hossein, Derayatifar Amir A, Curran Sarah
Department of Orthotics and Prosthetics, Faculty of Rehabilitation Sciences, Iran University of Medical sciences, Republic of Iran.
Department of Orthotics and Prosthetics, Faculty of Rehabilitation Sciences, Iran University of Medical sciences, Republic of Iran
Prosthet Orthot Int. 2015 Jun;39(3):190-6. doi: 10.1177/0309364614521652. Epub 2014 Mar 6.
Flexible flat foot is described as a reduction in the height of the medial longitudinal arch and may occur from abnormal foot pronation. A foot orthosis is thought to modify and control excessive pronation and improve arch height.
To compare the immediate effect of three types of orthoses on foot mobility and the arch height index in subjects with flexible flat feet.
A quasi-experimental study.
The dorsal arch height, midfoot width, foot mobility and arch height index were assessed in 20 participants with flexible flat feet (mean age = 23.2 ± 3 years) for three different foot orthosis conditions: soft, semi-rigid and rigid University of California Biomechanics Laboratory (UCBL).
Maximum midfoot width at 90% with arch mobility in the coronal plane was shown in the semi-rigid orthosis condition. The semi-rigid orthosis resulted in the highest mean foot mobility in 90% of weight bearing, and the rigid orthosis (UCBL) had the lowest mean foot mobility. The soft orthosis resulted in foot mobility between that of the rigid and the semi-rigid orthosis. UCBL orthosis showed the highest arch height index, and the semi-rigid orthosis showed the lowest mean arch height index.
Due to its rigid structure and long medial-lateral walls, the UCBL orthosis appears to limit foot mobility. Therefore, it is necessary to make an orthosis that facilitates foot mobility in the normal range of the foot arch. Future studies should address the dynamic mobility of the foot with using various types of foot orthoses.
Although there are many studies focussed on flat foot and the use of foot orthoses, the mechanism of action is still unclear. This study explored foot mobility and the influence of foot orthoses and showed that a more rigid foot orthosis should be selected based on foot mobility.
柔性扁平足被描述为内侧纵弓高度降低,可能由足部异常旋前引起。足部矫形器被认为可以改变和控制过度旋前,并提高足弓高度。
比较三种类型的矫形器对柔性扁平足受试者足部活动度和足弓高度指数的即时影响。
一项准实验研究。
对20名柔性扁平足受试者(平均年龄 = 23.2 ± 3岁)在三种不同的足部矫形器条件下评估足背弓高度、中足宽度、足部活动度和足弓高度指数:软质、半刚性和刚性的加利福尼亚大学生物力学实验室(UCBL)矫形器。
在半刚性矫形器条件下,在冠状面足弓活动度为90%时中足宽度最大。半刚性矫形器在90%负重时导致平均足部活动度最高,而刚性矫形器(UCBL)的平均足部活动度最低。软质矫形器导致的足部活动度介于刚性和半刚性矫形器之间。UCBL矫形器显示出最高的足弓高度指数,而半刚性矫形器显示出最低的平均足弓高度指数。
由于其刚性结构和长的内外侧壁,UCBL矫形器似乎会限制足部活动度。因此,有必要制作一种在足弓正常范围内促进足部活动度的矫形器。未来的研究应该使用各种类型的足部矫形器来研究足部的动态活动度。
尽管有许多研究关注扁平足和足部矫形器的使用,但其作用机制仍不清楚。本研究探讨了足部活动度以及足部矫形器的影响,并表明应根据足部活动度选择更刚性的足部矫形器。