Livingston A R, Callaghan F J, Byrd C L, Heemels J P, Hollander P, Van Mechelen R, Chappin J J
Telectronics and Cordis Pacing Systems, Miami, FL.
Pacing Clin Electrophysiol. 1988 Nov;11(11 Pt 2):1770-6. doi: 10.1111/j.1540-8159.1988.tb06308.x.
A new method of evoked response detection, previously demonstrated in the ventricle, has been studied in the atrium at the time of routine pacemaker implant in 16 patients. The atrial evoked response was readily detectable in all patients due to excellent recovery from poststimulus polarization. In six patients, as experimental threshold-tracking pacemaker was used to automatically verify atrial capture and to generate strength-duration curves. It is concluded that this pacing technique is both simple and reliable, and that automatic atrial threshold tracking is feasible.
一种先前已在心室中得到证实的诱发反应检测新方法,已在16例患者进行常规起搏器植入时在心房中进行了研究。由于刺激后极化恢复良好,所有患者的心房诱发反应均易于检测到。在6例患者中,使用实验性阈值跟踪起搏器自动验证心房捕获并生成强度-时间曲线。得出的结论是,这种起搏技术既简单又可靠,并且自动心房阈值跟踪是可行的。