Reducing image noise in computed tomography (CT) colonography: effect of an integrated circuit CT detector.

作者信息

Liu Yu, Leng Shuai, Michalak Gregory J, Vrieze Thomas J, Duan Xinhui, Qu Mingliang, Shiung Maria M, McCollough Cynthia H, Fletcher Joel G

机构信息

From the *Department of Radiology, Mayo Clinic, Rochester, MN; and †Siemens Medical Solutions, Malvern, PA.

出版信息

J Comput Assist Tomogr. 2014 May-Jun;38(3):398-403. doi: 10.1097/RCT.0000000000000081.

Abstract

OBJECTIVE

To investigate whether the integrated circuit (IC) detector results in reduced noise in computed tomography (CT) colonography (CTC).

METHODS

Three hundred sixty-six consecutive patients underwent clinically indicated CTC using the same CT scanner system, except for a difference in CT detectors (IC or conventional). Image noise, patient size, and scanner radiation output (volume CT dose index) were quantitatively compared between patient cohorts using each detector system, with separate comparisons for the abdomen and pelvis.

RESULTS

For the abdomen and pelvis, despite significantly larger patient sizes in the IC detector cohort (both P < 0.001), image noise was significantly lower (both P < 0.001), whereas volume CT dose index was unchanged (both P > 0.18). Based on the observed image noise reduction, radiation dose could alternatively be reduced by approximately 20% to result in similar levels of image noise.

CONCLUSION

Computed tomography colonography images acquired using the IC detector had significantly lower noise than images acquired using the conventional detector. This noise reduction can permit further radiation dose reduction in CTC.

摘要

目的

研究集成电路(IC)探测器是否能降低计算机断层扫描(CT)结肠成像(CTC)中的噪声。

方法

366例连续患者接受了临床指征的CTC检查,使用的是同一台CT扫描仪系统,只是CT探测器不同(IC探测器或传统探测器)。使用每种探测器系统,对患者队列的图像噪声、患者体型和扫描仪辐射输出(容积CT剂量指数)进行定量比较,并分别对腹部和骨盆进行比较。

结果

对于腹部和骨盆,尽管IC探测器队列中的患者体型明显更大(P均<0.001),但图像噪声明显更低(P均<0.001),而容积CT剂量指数未变(P均>0.18)。基于观察到的图像噪声降低情况,辐射剂量可减少约20%,以产生相似水平的图像噪声。

结论

使用IC探测器获取的CT结肠成像图像的噪声明显低于使用传统探测器获取的图像。这种噪声降低可以在CTC中进一步降低辐射剂量。