Photon counting detector (PCD)-based computed tomography (CT) is an emerging imaging technique. Compared to conventional energy integrating detector (EID)-based CT, PCD-CT is able to exclude electronic noise that may severely impair image quality at low photon counts. This work focused on comparing the noise performance at low doses between the PCD and EID subsystems of a whole-body research PCD-CT scanner, both qualitatively and quantitatively. An anthropomorphic thorax phantom was scanned, and images of the shoulder portion were reconstructed. The images were visually and quantitatively compared between the two subsystems in terms of streak artifacts, an indicator of the impact of electronic noise. Furthermore, a torso-shaped water phantom was scanned using a range of tube currents. The product of the noise and the square root of the tube current was calculated, normalized, and compared between the EID and PCD subsystems. Visual assessment of the thorax phantom showed that electronic noise had a noticeably stronger degrading impact in the EID images than in the PCD images. The quantitative results indicated that in low-dose situations, electronic noise had a noticeable impact (up to a 5.8% increase in magnitude relative to quantum noise) on the EID images, but negligible impact on the PCD images.
基于光子计数探测器(PCD)的计算机断层扫描(CT)是一种新兴的成像技术。与传统的基于能量积分探测器(EID)的CT相比,PCD-CT能够排除在低光子计数时可能严重损害图像质量的电子噪声。这项工作专注于定性和定量地比较全身研究型PCD-CT扫描仪的PCD和EID子系统在低剂量下的噪声性能。扫描了一个仿真人体胸部模型,并重建了肩部部分的图像。从条纹伪影(电子噪声影响的一个指标)方面对两个子系统的图像进行了视觉和定量比较。此外,使用一系列管电流扫描了一个躯干形状的水体模。计算、归一化并比较了EID和PCD子系统之间噪声与管电流平方根的乘积。对胸部模型的视觉评估表明,电子噪声在EID图像中的降解影响明显强于PCD图像。定量结果表明,在低剂量情况下,电子噪声对EID图像有显著影响(相对于量子噪声,幅度增加高达5.8%),但对PCD图像的影响可忽略不计。