Suppr超能文献

采用DMEQ-TAD衍生化的液相色谱-串联质谱法同时定量测定25-羟基维生素D和24,25-二羟基维生素D的临床应用

Clinical utility of simultaneous quantitation of 25-hydroxyvitamin D and 24,25-dihydroxyvitamin D by LC-MS/MS involving derivatization with DMEQ-TAD.

作者信息

Kaufmann Martin, Gallagher J Christopher, Peacock Munro, Schlingmann Karl-Peter, Konrad Martin, DeLuca Hector F, Sigueiro Rita, Lopez Borja, Mourino Antonio, Maestro Miguel, St-Arnaud René, Finkelstein Joel S, Cooper Donald P, Jones Glenville

机构信息

Department of Biomedical and Molecular Sciences (M.Ka., G.J.), Queen's University, Kingston, Ontario, Canada K7L 3N6; Bone Metabolism Unit (J.C.G.), Creighton University School of Medicine, Omaha, Nebraska 68131; Indiana University School of Medicine (M.P.), Indianapolis, Indiana 46202-5250; Department of General Pediatrics, (K-P. S., M.Ko.) University Children's Hospital, D-48149, Münster, Germany; Department of Biochemistry (H.F.D.), University of Wisconsin-Madison, Madison, Wisconsin 53706; Departmento de Quimica Organica (R.S., B.L., A.M., M.M.), Laboratorio Ignacio Ribas, University of Santiago de Compostela, 15782 Santiago de Compostela, Spain; McGill University and Shriners Hospital for Children-Canada (R.S.), Montréal, Québec, Canada H3G 1A6; Endocrine Unit (J.S.F.), Massachusetts General Hospital, Boston Massachusetts 02114; and Health Sciences Research, Waters Corporation (D.P.C), Stamford Avenue, Altrincham Rd, Wilmslow, SK9 4AX United Kingdom.

出版信息

J Clin Endocrinol Metab. 2014 Jul;99(7):2567-74. doi: 10.1210/jc.2013-4388. Epub 2014 Mar 26.

Abstract

CONTEXT

The discovery of hypercalcemic diseases due to loss-of-function mutations in 25-hydroxyvitamin D-24-hydroxylase has placed a new demand for sensitive and precise assays for 24,25-dihydroxyvitamin D [24,25-(OH)2D].

OBJECTIVE

We describe a novel liquid chromatography and tandem mass spectrometry-based method involving derivatization with DMEQ-TAD {4-[2-(6,7-dimethoxy-4-methyl-3,4-dihydroquinoxalinyl)ethyl]-1,2,4-triazoline-3,5-dione} to simultaneously assay multiple vitamin D metabolites including 25-hydroxyvitamin D (25-OH-D) and 24,25-(OH)2D using 100 μL of serum with a 5-minute run time.

DESIGN

The assay uses a newly synthesized internal standard d6-24,25-(OH)2D3 enabling the quantitation of 24,25-(OH)2D3 as well as the determination of the ratio of 25-OH-D3 to 24,25-(OH)2D3, a physiologically useful parameter.

SETTING

We report data on more than 1000 normal and disease samples involving vitamin D deficiency or hypercalcemia in addition to studies involving knockout mouse models.

RESULTS

The assay showed good correlation with samples from quality assurance schemes for 25-OH-D (25-OH-D2 and 25-OH-D3) determination (-2% to -5% bias) and exhibited low inter- and intraassay coefficients of variation (4%-7%) and lower limits of quantitation of 0.25-0.45 nmol/L. In clinical studies, we found a strong correlation between serum levels of 25-OH-D3 and 24,25-(OH)2D3 (r(2) = 0.80) in subjects over a broad range of 25-OH-D3 values and a marked lack of production of 24,25-(OH)2D3 below 25 nmol/L of 25-OH-D. The ratio of 25-OH-D3 to 24,25-(OH)2D3, which remained less than 25 in vitamin D-sufficient subjects (serum 25-OH-D < 50 nmol/L) but was greatly elevated (80-100) in patients with idiopathic infantile hypercalcemia.

CONCLUSIONS

The new method showed good utility in clinical settings involving vitamin D deficiency; supplementation with vitamin D and idiopathic infantile hypercalcemia, as well as in animal models with ablation of selected cytochrome P450-containing enzymes involved in vitamin D metabolism.

摘要

背景

因25-羟基维生素D-24-羟化酶功能丧失突变导致的高钙血症疾病的发现,对24,25-二羟基维生素D[24,25-(OH)₂D]的灵敏且精确的检测方法提出了新的要求。

目的

我们描述了一种基于液相色谱和串联质谱的新方法,该方法涉及用DMEQ-TAD{4-[2-(6,7-二甲氧基-4-甲基-3,4-二氢喹喔啉基)乙基]-1,2,4-三唑啉-3,5-二酮}进行衍生化,以使用100μL血清并在5分钟的运行时间内同时检测多种维生素D代谢物,包括25-羟基维生素D(25-OH-D)和24,25-(OH)₂D。

设计

该检测方法使用新合成的内标d6-24,25-(OH)₂D₃,能够对24,25-(OH)₂D₃进行定量,并确定25-OH-D₃与24,25-(OH)₂D₃的比值,这是一个具有生理意义的参数。

设置

除了涉及基因敲除小鼠模型的研究外,我们报告了超过1000份涉及维生素D缺乏或高钙血症的正常和疾病样本的数据。

结果

该检测方法与用于25-OH-D(25-OH-D₂和25-OH-D₃)测定的质量保证方案中的样本具有良好的相关性(偏差为-2%至-5%),并且具有较低的批间和批内变异系数(4%-7%)以及0.25-0.45nmol/L的定量下限。在临床研究中,我们发现,在25-OH-D₃值范围广泛的受试者中,血清25-OH-D₃水平与24,25-(OH)₂D₃水平之间存在很强的相关性(r² = 0.80),并且当25-OH-D低于25nmol/L时,24,25-(OH)₂D₃的生成明显缺乏。在维生素D充足的受试者(血清25-OH-D < 50nmol/L)中,25-OH-D₃与24,25-(OH)₂D₃的比值保持小于25,但在特发性婴儿高钙血症患者中该比值大幅升高(80-100)。

结论

这种新方法在涉及维生素D缺乏、维生素D补充和特发性婴儿高钙血症的临床环境中,以及在涉及维生素D代谢中特定含细胞色素P450酶缺失的动物模型中显示出良好的实用性。

相似文献

引用本文的文献

本文引用的文献

5
Effects of vitamin D supplementation in older African American women.维生素 D 补充对老年非裔美国女性的影响。
J Clin Endocrinol Metab. 2013 Mar;98(3):1137-46. doi: 10.1210/jc.2012-3106. Epub 2013 Feb 5.
10
Metabolism and biomarkers of vitamin D.维生素D的代谢与生物标志物
Scand J Clin Lab Invest Suppl. 2012;243:7-13. doi: 10.3109/00365513.2012.681892.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验