Suppr超能文献

使用在板与杆之间带有球窝关节的部分听骨置换假体的经验。

Experience with the use of a partial ossicular replacement prosthesis with a ball-and-socket joint between the plate and the shaft.

作者信息

Birk Stephanie, Brase Christoph, Hornung Joachim

机构信息

Department of Ear, Nose and Throat, Head and Neck - Surgery, University Hospital of Erlangen, Friedrich-Alexander University Erlangen-Nürnberg, Erlangen, Germany.

出版信息

Otol Neurotol. 2014 Aug;35(7):1248-50. doi: 10.1097/MAO.0000000000000383.

Abstract

BACKGROUND

In the further development of alloplastic prostheses for use in middle ear surgery, the Dresden and Cologne University Hospitals, working together with a company, introduced a new partial ossicular replacement prosthesis in 2011. The ball-and-socket joint between the prosthesis and the shaft mimics the natural articulations between the malleus and incus and between the incus and stapes, allowing reaction to movements of the tympanic membrane graft, particularly during the healing process.

STUDY DESIGN

Retrospective evaluation

METHODS

To reconstruct sound conduction as part of a type III tympanoplasty, partial ossicular replacement prosthesis with a ball-and-socket joint between the plate and the shaft was implanted in 60 patients, with other standard partial ossicular replacement prosthesis implanted in 40 patients and 64 patients. Pure-tone audiometry was carried out, on average, 19 and 213 days after surgery. Results of the partial ossicular replacement prosthesis with a ball-and-socket joint between the plate and the shaft were compared with those of the standard prostheses.

RESULTS

Early measurements showed a mean improvement of 3.3 dB in the air-bone gap (ABG) with the partial ossicular replacement prosthesis with a ball-and-socket joint between the plate and the shaft, giving similar results than the standard implants (6.6 and 6.0 dB, respectively), but the differences were not statistically significant. Later measurements showed a statistically significant improvement in the mean ABG, 11.5 dB, compared with 4.4 dB for one of the standard partial ossicular replacement prosthesis and a tendency of better results to 6.9 dB of the other standard prosthesis.

CONCLUSIONS

In our patients, we achieved similarly good audiometric results to those already published for the partial ossicular replacement prosthesis with a ball-and-socket joint between the plate and the shaft. Intraoperative fixation posed no problems, and the postoperative complication rate was low.

摘要

背景

在中耳手术用异体植入假体的进一步研发过程中,德累斯顿大学医院和科隆大学医院与一家公司合作,于2011年推出了一种新型部分听骨链置换假体。该假体与柄部之间的球窝关节模拟了锤骨与砧骨之间以及砧骨与镫骨之间的自然关节,能够对鼓膜移植物的运动做出反应,尤其是在愈合过程中。

研究设计

回顾性评估

方法

作为III型鼓室成形术的一部分,为重建声音传导,60例患者植入了板与柄部之间带有球窝关节的部分听骨链置换假体,40例和64例患者分别植入了其他标准部分听骨链置换假体。术后平均19天和213天进行纯音听力测定。将板与柄部之间带有球窝关节的部分听骨链置换假体的结果与标准假体的结果进行比较。

结果

早期测量显示,板与柄部之间带有球窝关节的部分听骨链置换假体使气骨导差(ABG)平均改善3.3 dB,与标准植入物的结果相似(分别为6.6 dB和6.0 dB),但差异无统计学意义。后期测量显示,平均ABG有统计学意义的改善,为11.5 dB,而其中一种标准部分听骨链置换假体为4.4 dB,另一种标准假体为6.9 dB,有更好结果的趋势。

结论

在我们的患者中,我们获得了与已发表的板与柄部之间带有球窝关节的部分听骨链置换假体类似良好的听力测定结果。术中固定没有问题,术后并发症发生率低。