灭菌和使用次数对牙科种植体摩擦式机械扭矩限制装置准确性的影响。

The effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants.

作者信息

Fayaz Ali, Mahshid Minoo, Saboury Aboulfazl, Sadr Seyed Jalil, Ansari Ghassem

机构信息

Department of Prosthodontics, Dental School, Shahid Beheshti University of Medical Sciences, Tehran, Iran.

Department of Pedodontics, Dental School, Shahid Beheshti University of Medical Sciences, Tehran, Iran.

出版信息

Dent Res J (Isfahan). 2014 Jan;11(1):74-80.

Abstract

BACKGROUND

Mechanical torque limiting devices (MTLDs) are necessary tools to control a peak torque and achieving target values of screw component of dental implants. Due to probable effect of autoclaving and number of use on the accuracy of these devices, this study aimed to evaluate the effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices (F-S MTLDs) in achieving their target torque values.

MATERIALS AND METHODS

Peak torque measurements of 15 new F-S MTLDs from three different manufacturers (Astra Tech, BioHorizons, Dr. Idhe) were measured ten times before and after 100 steam sterilization using a digital torque gauge. To simulate the clinical situation of aging (number of use) target torque application process was repeated 10 times after each sterilization cycle and the peak torque values were registered. Comparison of the mean differences with target torque in each cycle was performed using one sample t test. Considering the type of MTLDs as inter subject comparison, One-way repeated measure ANOVA was used to evaluate the absolute values of differences between devices of each manufacturer in each group (α = 0.05).

RESULTS

The results of this study in Dr. Idhe group showed that, mean of difference values significantly differed from the target torque (P = 0.002) until 75 cycles. In Astra Tech group, also mean of difference values with under estimation trend, showed a significant difference with the target torque (P < 0.001). Mean of difference values significantly differed from the target torque with under estimation trend during all the 100 cycles in BioHorizons group (P < 0.05).

CONCLUSION

The torque output of each individual device stayed in 10% difference from target torque values before 100 sterilization cycles, but more than 10% difference from the target torque was seen in varying degrees during these consequent cycles.

摘要

背景

机械扭矩限制装置(MTLDs)是控制峰值扭矩并实现牙种植体螺钉部件目标值的必要工具。由于高压灭菌和使用次数可能会对这些装置的准确性产生影响,本研究旨在评估灭菌和使用次数对摩擦式机械扭矩限制装置(F-S MTLDs)实现其目标扭矩值准确性的影响。

材料与方法

使用数字扭矩计,对来自三个不同制造商(Astra Tech、BioHorizons、Dr. Idhe)的15个新型F-S MTLDs在100次蒸汽灭菌前后各测量峰值扭矩10次。为模拟老化(使用次数)的临床情况,在每个灭菌周期后重复目标扭矩施加过程10次,并记录峰值扭矩值。使用单样本t检验比较每个周期中与目标扭矩的平均差异。将MTLDs的类型作为受试者间比较,采用单向重复测量方差分析来评估每组中每个制造商的装置之间差异的绝对值(α = 0.05)。

结果

本研究在Dr. Idhe组的结果显示,直到75个周期,差异值的平均值与目标扭矩有显著差异(P = 0.002)。在Astra Tech组中,差异值的平均值也呈低估趋势,与目标扭矩有显著差异(P < 0.001)。在BioHorizons组的所有100个周期中,差异值的平均值呈低估趋势,与目标扭矩有显著差异(P < 0.05)。

结论

在100次灭菌周期之前,每个装置的扭矩输出与目标扭矩值的差异保持在10%以内,但在随后的这些周期中,不同程度地出现了与目标扭矩超过10%的差异。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/f677/3955319/aa1271d0b5f1/DRJ-11-74-g001.jpg