J Biomech. 2014 Jun 3;47(8):1914-7. doi: 10.1016/j.jbiomech.2014.03.034. Epub 2014 Apr 3.
A repeatable method for in vivo and in vitro measurement of polyethylene wear in total knee replacement (TKA) is needed. This research examines the model-based radiostereometric analysis' (MBRSA) in vitro precision under different patient-radiograph orientations and flexion angles of the knee using a TKA phantom. Anterior-posterior and medial-lateral imaging orientations showed the highest precision; better than 0.036mm (3-dimensional translation) and 0.089° (3-dimensional rotation). Flexion of the knee did not affect MBRSA precision. Medial-lateral imaging is advantageous as it allows for flexion of the knee joint during an RSA examination, thus providing greater information for wear measurement.