Sariali E, Klouche S, Mamoudy P
Service d'orthopédie traumatologique, hôpital Pitié-Salpétrière, 83, boulevard de l'Hôpital, 75013 Paris, France.
Service d'orthopédie traumatologique, hôpital Pitié-Salpétrière, 83, boulevard de l'Hôpital, 75013 Paris, France.
Orthop Traumatol Surg Res. 2014 Jun;100(4):437-40. doi: 10.1016/j.otsr.2014.01.009. Epub 2014 Apr 13.
Many factors were incriminated in the squeaking generation in ceramic-on-ceramic total hip arthroplasty (THA), including the cup positioning and design. However, the influence of the stem orientation has not been investigated and the true three-dimensional hip anatomy has never been compared to the contralateral healthy hip. Three patients, who underwent unilateral ceramic-on-ceramic THA, complained of squeaking. CT-scans were performed to compare the true three-dimensional hip anatomy to the contralateral healthy hip. All patients presented evidence of posterior neck-rim impingement with a two-fold increase in the global anteversion (above 75°) comparatively to the healthy hip. The excess of anteversion was on the cup side in 2 cases and on the stem side in 1 case. We conclude that squeaking in ceramic-on-ceramic THA could be related to a poor accuracy of 3D hip anatomy reconstruction which generated a posterior impingement and subsequent anterior edge loading because of excessive global anteversion.
在陶瓷对陶瓷全髋关节置换术(THA)中,有许多因素被认为与摩擦音的产生有关,包括髋臼杯的位置和设计。然而,股骨柄方向的影响尚未得到研究,而且从未将真实的三维髋关节解剖结构与对侧健康髋关节进行比较。三名接受单侧陶瓷对陶瓷THA的患者抱怨出现了摩擦音。进行了CT扫描,以将真实的三维髋关节解剖结构与对侧健康髋关节进行比较。所有患者均表现出后颈部-髋臼边缘撞击的迹象,与健康髋关节相比,整体前倾角增加了两倍(超过75°)。前倾角过大在2例中出现在髋臼杯一侧,1例出现在股骨柄一侧。我们得出结论,陶瓷对陶瓷THA中的摩擦音可能与三维髋关节解剖结构重建的准确性差有关,这会导致后撞击以及由于整体前倾角过大而导致的前缘负荷。