Flamant Fabrice, Borg Jeanne-Yvonne, Lenormand Bernard, Le Cam-Duchez Véronique
Unité fonctionnelle d'hémostase, hématologie biologique, Institut de biologie clinique, CHU de Rouen, France.
Ann Biol Clin (Paris). 2014 Mar-Apr;72(2):231-5. doi: 10.1684/abc.2014.0941.
Actually, many laboratories tend to acquire pre analytic automates to prepare specimens for analysis. For haemostasis, these pre analytical modules are not always in agreement with the recommendations from the Groupe d'étude de l'hémostase et de la thrombose (GEHT). For example in the MPA C10 module (Roche Diagnostics) the speed of centrifugation was not rather fast compared with the GEHT recommandations. Then, to be able to use this automate for routine coagulation assays, we compared results of Quick time, activated partial prothombin time, fibrinogen, factor II, factor V, factor VII, factor X and antithrombin levels and unfractioned heparin anti-Xa activity measurement after MPA (1,885 g - 999 sec) or GEHT (2,500 g - 900 sec) protocol of centrifugation. First, we verified platelet counts: in 82% of specimens, the platelet counts were under 10.10(9)/L after centrifugation on MPA module. Moreover, a good correlation was observed in all comparisons. Then we concluded the MPA C10 module was usable for routine coagulation tests.