Gamma 3 与 Intertan 钉系统治疗股骨转子间骨折时的拉力螺钉位置不良:一项生物力学比较。

Malpositioning of the lag screws by 1- or 2-screw nailing systems for pertrochanteric femoral fractures: a biomechanical comparison of gamma 3 and intertan.

机构信息

*Department of Trauma, Hand and Reconstructive Surgery, University Medical Center Hamburg-Eppendorf, Hamburg, Germany; †Biomechanics Section, Hamburg University of Technology, Hamburg, Germany; and ‡Department of Forensic Medicine, University Medical Center Hamburg-Eppendorf, Hamburg, Germany.

出版信息

J Orthop Trauma. 2014 May;28(5):276-82. doi: 10.1097/BOT.0000000000000008.

Abstract

OBJECTIVES

The aim of this investigation was to perform a biomechanical comparison between 1- and 2-screw systems used for the treatment of intertrochanteric fractures for centralized and decentralized placement of femoral neck screws of failure loads, stiffness, survival rates, tip apex distance (TAD), and failure mode.

METHODS

As fracture model, an AO 31A2.3 fracture was used. Twelve pairs of human cadaver femora were tested. Femoral neck screws were implanted in the femoral head in center/center, posterior/central, and anterior/superior position in axial/frontal plane. A single-screw system (Gamma 3 Locking Nail; Stryker GmbH & Co. KG) and a 2-screw system (Trigen-Intertan; Smith & Nephew GmbH) were used. To simulate the load in situ, a cyclic load was carried for 10,000 cycles in a material testing machine. If no cyclic failure occurred, femora were loaded until the failure. The systems were compared according to the stiffness, survivability through 10 k cycles, TAD, and load to failure.

RESULTS

None of the tested bones failed at center/center location in the decentralized positions 3 Gamma Nail and 2 Intertan specimens failed during cyclic testing. The 2-screw system resisted higher forces in all positions (Gamma: 5370N ± 1924, Intertan: 7650N ± 2043; P = 0.014).

CONCLUSIONS

Based on these data, it is clear that both the nail systems showed a higher biomechanical stability with a lower TAD. The 2 specimens that failed with the Intertan in the cyclic tests had a TAD ≥49 mm. The cutout failures that we detected during cyclic testing in the Gamma system had a TAD ≥30 mm. Thus, it is clear that the TAD affects failure independent of the implant used. With a less than ideal lag screw placement, however, the Intertan system with 2 integrated screws was able to withstand higher loads in this study.

摘要

目的

本研究旨在对用于治疗股骨转子间骨折的 1 螺钉和 2 螺钉系统进行生物力学比较,比较集中和分散放置股骨颈螺钉后的失效负荷、刚度、存活率、尖端顶点距离(TAD)和失效模式。

方法

采用 AO 31A2.3 骨折模型,对 12 对人体股骨标本进行测试。在轴向/额状面,将股骨颈螺钉植入股骨头的中心/中心、后/中及前/上位置。使用单螺钉系统(Gamma 3 锁定钉;Stryker GmbH & Co. KG)和双螺钉系统(Trigen-Intertan;Smith & Nephew GmbH)。为了模拟原位载荷,在材料试验机上进行了 10000 次循环加载。如果没有发生循环失效,则对股骨进行加载,直到失效。根据刚度、通过 10 k 次循环的存活率、TAD 和失效负荷对系统进行比较。

结果

在分散位置 3 处,Gamma 钉和 2 个 Intertan 标本在中心/中心位置的所有测试骨均未失效,在循环测试中失败。在所有位置,双螺钉系统均能抵抗更高的力(Gamma:5370N±1924,Intertan:7650N±2043;P=0.014)。

结论

根据这些数据,可以清楚地看出,两种钉系统的 TAD 较低,生物力学稳定性更高。在循环测试中与 Intertan 一起失效的 2 个标本的 TAD≥49mm。在 Gamma 系统的循环测试中检测到的切割失效的 TAD≥30mm。因此,很明显,TAD 会影响失效,而与使用的植入物无关。然而,在不理想的拉力螺钉放置情况下,在这项研究中,具有 2 个集成螺钉的 Intertan 系统能够承受更高的载荷。