Dierckx D, Saldarriaga Vargas C, Rogge F, Lichtherte S, Struelens L
Department of Medical Physics, Controlatom, Jan Olieslagerslaan 35, 1800 Vilvoorde, Belgium
SCK-CEN, Radiation Protection, Dosimetry and Calibration, Boeretang 200, 2400 Mol, Belgium.
Radiat Prot Dosimetry. 2015 Jan;163(1):125-32. doi: 10.1093/rpd/ncu117. Epub 2014 Apr 24.
The use of cone beam computed tomography (CBCT) in diagnostic radiology departments is increasing. Several discussions arise whether with the CBCT application, some multi-slice CT (MSCT) examinations can be replaced by it. High hopes are set regarding the dosimetric aspects of CBCT: are patient doses in between those of conventional X-rays and MSCT? In this study, effective dose and organ doses were evaluated for two non-dental CBCT examinations: sinus and middle ear. A comparison with the dose obtained with a MSCT protocol was performed. Moreover, the sinus examination was also compared with the dose obtained by projection radiography (RX). Effective doses were estimated from thermoluminescent detector dose measurements in an anthropomorphic phantom and were compared against Monte Carlo simulations. Results show that the effective dose for the sinus examination is more than three times higher with MSCT than with CBCT and about five times lower with RX compared with CBCT, whereas for the middle ear examination, the effective dose obtained with MSCT is almost six times higher than that of CBCT. Finally, a sensitivity study on the size and position of the CBCT field of view showed the influence of these two factors on the dose received by the patient.
锥形束计算机断层扫描(CBCT)在诊断放射科的使用正在增加。关于CBCT的应用是否可以取代一些多层螺旋CT(MSCT)检查,出现了一些讨论。人们对CBCT的剂量学方面寄予厚望:患者剂量是否介于传统X射线和MSCT之间?在本研究中,对两种非牙科CBCT检查(鼻窦和中耳)的有效剂量和器官剂量进行了评估。并与MSCT方案获得的剂量进行了比较。此外,鼻窦检查还与投影放射成像(RX)获得的剂量进行了比较。有效剂量通过在人体模型中进行热释光探测器剂量测量来估计,并与蒙特卡罗模拟结果进行比较。结果表明,鼻窦检查的有效剂量,MSCT比CBCT高出三倍多,RX比CBCT低约五倍;而对于中耳检查,MSCT获得的有效剂量几乎比CBCT高六倍。最后,对CBCT视野大小和位置的敏感性研究表明了这两个因素对患者接受剂量的影响。