Seiden-Long Isolde, Schnabl Kareena, Skoropadyk Wendy, Lennon Nola, McKeage Arlayne
Calgary Laboratory Services, Calgary, AB, Canada; Department of Pathology and Laboratory Medicine, University of Calgary, Canada.
Department of Laboratory Medicine and Pathology, University of Alberta, Canada.
Clin Biochem. 2014 Aug;47(12):1116-20. doi: 10.1016/j.clinbiochem.2014.04.022. Epub 2014 May 2.
Adaptation of the Randox Enzymatic Manual UV Ammonia method to be used on the Roche Cobas 6000 (c501) automated analyzer platform.
The Randox ammonia reagent was evaluated for precision, linearity, accuracy and interference from hemolysis, icterus and lipemia on the Roche c501 analyzer. Comparison studies were conducted for the Randox reagent between Roche c501, Siemens Vista, Ortho Vitros 250, and Beckman DxC methods.
The Randox reagent demonstrates acceptable within-run (L1=65 μmol/L, CV 3.4% L2=168 μmol/L, CV 1.9%) and between-run precision (L1=29 μmol/L, CV 7.3% L2=102 μmol/L, CV 3.0%), Analytical Measurement Range (7-940 μmol/L), and accuracy. The method interference profile is superior for the Randox method (hemolysis index up to 600, icteric index up to 60, lipemic index up to 100) as compared to the Roche method (hemolysis index up to 200, icteric index up to 10, lipemic index up to 50). Comparison was very good between the Randox reagent and two other wet chemistry platforms.
The Randox Enzymatic Manual UV Ammonia reagent is an available alternative to the Roche Cobas c501 reagent. The method is more robust to endogenous interferences and less prone to instrument error flags, thus allowing the majority of clinical specimens to be reported without additional sample handling at our institution.
使兰多克斯酶法手动紫外氨测定法适用于罗氏Cobas 6000(c501)自动化分析仪平台。
在罗氏c501分析仪上对兰多克斯氨试剂的精密度、线性、准确性以及溶血、黄疸和脂血干扰进行评估。对罗氏c501、西门子Vista、奥多维他斯250和贝克曼DxC方法之间的兰多克斯试剂进行了比较研究。
兰多克斯试剂显示出可接受的批内精密度(L1 = 65 μmol/L,CV 3.4%;L2 = 168 μmol/L,CV 1.9%)和批间精密度(L1 = 29 μmol/L,CV 7.3%;L2 = 102 μmol/L,CV 3.0%)、分析测量范围(7 - 940 μmol/L)以及准确性。与罗氏方法(溶血指数高达200,黄疸指数高达10,脂血指数高达50)相比,兰多克斯方法的方法干扰谱更优(溶血指数高达600,黄疸指数高达60,脂血指数高达100)。兰多克斯试剂与其他两个湿化学平台之间的比较结果非常好。
兰多克斯酶法手动紫外氨试剂是罗氏Cobas c501试剂的一种可行替代方案。该方法对内源性干扰更具耐受性,且较少出现仪器错误提示,因此在我们机构,大多数临床标本无需额外的样本处理即可报告结果。