Effect of cup inclination on predicted contact stress-induced volumetric wear in total hip replacement.

作者信息

Rijavec B, Košak R, Daniel M, Kralj-Iglič V, Dolinar D

机构信息

a Laboratory of Clinical Biophysics, Faculty of Health Sciences , University of Ljubljana , Zdravstvena 5, SI-1000 Ljubljana , Slovenia.

出版信息

Comput Methods Biomech Biomed Engin. 2015;18(13):1468-73. doi: 10.1080/10255842.2014.916700. Epub 2014 May 16.

Abstract

In order to increase the lifetime of the total hip endoprosthesis, it is necessary to understand mechanisms leading to its failure. In this work, we address volumetric wear of the artificial cup, in particular the effect of its inclination with respect to the vertical. Volumetric wear was calculated by using mathematical models for resultant hip force, contact stress and penetration of the prosthesis head into the cup. Relevance of the dependence of volumetric wear on inclination of the cup (its abduction angle ϑA) was assessed by the results of 95 hips with implanted endoprosthesis. Geometrical parameters obtained from standard antero-posterior radiographs were taken as input data. Volumetric wear decreases with increasing cup abduction angle ϑA. The correlation within the population of 95 hips was statistically significant (P = 0.006). Large cup abduction angle minimises predicted volumetric wear but may increase the risk for dislocation of the artificial head from the cup in the one-legged stance. Cup abduction angle and direction of the resultant hip force may compensate each other to achieve optimal position of the cup with respect to wear and dislocation in the one-legged stance for a particular patient.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验