用于早期青光眼检测的激光扫描诊断设备的比较

Comparison of Laser Scanning Diagnostic Devices for Early Glaucoma Detection.

作者信息

Schulze Andreas, Lamparter Julia, Pfeiffer Norbert, Berisha Fatmire, Schmidtmann Irene, Hoffmann Esther M

机构信息

Departments of *Ophthalmology †Medical Biometry, Epidemiology and Informatics (IMBEI), University Medical Center Mainz, Mainz, Germany.

出版信息

J Glaucoma. 2015 Aug;24(6):442-7. doi: 10.1097/IJG.0000000000000054.

Abstract

BACKGROUND

To compare the diagnostic accuracy and to evaluate the correlation of optic nerve head and retinal nerve fiber layer thickness values between Fourier-Domain optical coherence tomography (FD-OCT), confocal scanning laser ophthalmoscopy (CSLO), and scanning laser polarimetry (SLP) for early glaucoma detection.

PATIENTS AND METHODS

Ninety-three patients with early open-angle glaucoma, 58 patients with ocular hypertension, and 60 healthy control subjects were included in this observational, cross-sectional study. All study participants underwent FD-OCT (RTVue-100), CSLO (HRT3), and SLP (GDx VCC) imaging of the optic nerve head and the retinal nerve fiber layer. Area under the receiver operating characteristic curves (AUROC) and Bland-Altman analysis were performed.

RESULTS

The parameters with the highest diagnostic accuracy were found for FD-OCT cup-to-disc ratio (AUROC=0.841), for SLP NFI (AUROC=0.835), and for CSLO cup-to-disc ratio (AUROC=0.789). Diagnostic accuracy of the best CSLO and SLP parameter was similar (P=0.259). There was a small statistically significant difference between the best CSLO and FD-OCT parameters for differentiating between glaucoma and healthy eyes (P=0.047).

CONCLUSIONS

FD-OCT and SLP have a similarly good diagnostic ability to distinguish between early glaucoma and healthy subjects. The diagnostic accuracy of CSLO was comparable with SLP and marginally lower compared with FD-OCT.

摘要

背景

比较傅里叶域光学相干断层扫描(FD-OCT)、共焦扫描激光检眼镜(CSLO)和扫描激光偏振仪(SLP)在早期青光眼检测中对视神经乳头和视网膜神经纤维层厚度值的诊断准确性,并评估其相关性。

患者与方法

这项观察性横断面研究纳入了93例早期开角型青光眼患者、58例高眼压症患者和60例健康对照者。所有研究参与者均接受了视神经乳头和视网膜神经纤维层的FD-OCT(RTVue-100)、CSLO(HRT3)和SLP(GDx VCC)成像。进行了受试者操作特征曲线下面积(AUROC)分析和布兰德-奥特曼分析。

结果

FD-OCT杯盘比(AUROC=0.841)、SLP神经纤维指数(AUROC=0.835)和CSLO杯盘比(AUROC=0.789)的诊断准确性最高。CSLO和SLP最佳参数的诊断准确性相似(P=0.259)。在区分青光眼和健康眼方面,CSLO和FD-OCT最佳参数之间存在微小的统计学显著差异(P=0.047)。

结论

FD-OCT和SLP在区分早期青光眼和健康受试者方面具有相似的良好诊断能力。CSLO的诊断准确性与SLP相当,与FD-OCT相比略低。