Mercer's Institute for Research on Ageing, St. James's Hospital, Dublin 8, Ireland.
Physiol Meas. 2014 Jul;35(7):1229-43. doi: 10.1088/0967-3334/35/7/1229. Epub 2014 May 22.
Ocular microtremor (OMT) is a small involuntary eye movement present in all subjects. In this paper we present the results of in vivo OMT measurement using a novel non-contact laser speckle technique. OMT signals have not previously been measured from the sclera using this laser speckle correlation technique. To verify the system's ability to record eye movements, it is first tested using a large angle eye rotation. Next, the system is tested with a group of 20 subjects and OMT parameters are extracted. The results of OMT measurements gave a mean frequency of 78 ± 3.86 Hz and peak-to-peak amplitude of 21.42 ± 7.01 µrad, these values are consistent with known values from eye-contacting methods.
眼震(OMT)是一种存在于所有受试者中的微小的无意识眼动。本文采用一种新的非接触激光散斑技术,介绍了眼震的活体测量结果。以前从未使用这种激光散斑相关技术从巩膜测量眼震信号。为了验证系统记录眼球运动的能力,首先使用大角度眼球旋转对系统进行测试。然后,用 20 名受试者对系统进行测试,并提取眼震参数。眼震测量结果的平均值为 78 ± 3.86 Hz,峰峰值为 21.42 ± 7.01 µrad,这些值与接触眼球的方法得出的已知值一致。