Alkiewicz J, Kedzia B, Hołderna E, Bugaj U
Pol Tyg Lek. 1989;44(34-35):813-6.
It was found that inflating tube is most rarely contaminated with microorganisms during the use of ultrasonic inhalator TUR USI 70. Glass cylinder is contaminated more frequently whereas a diaphragm, aerosol preparation, inhaling mask and a pipe joining it with the device are contaminated most frequently. Sporadic contamination of the inflating tube indicate an efficient work of air filters while frequent contamination of the diaphragm, aerosol preparation and glass cylinder prove that the contamination is caused by a coupling fluid. It was also found that ultrasound exerts a destructive effect on microorganisms in the aerosol preparation. The investigations have shown that the inhaling mask and tubes joining it with the device should be changed before each use while the other parts of an inhalator and aerosol preparation may be changed once per 15 inhalations. It was also noted that disinfection of different parts of the device by a 2% aqueous glutaric aldehyde (30 minutes at room temperature) is efficient in about 95%.
研究发现,在使用超声雾化器TUR USI 70的过程中,充气管受微生物污染的情况最为罕见。玻璃圆筒受污染的频率更高,而隔膜、气雾剂制剂、吸入面罩以及将其与设备相连的管道受污染的频率最高。充气管的偶发污染表明空气过滤器工作有效,而隔膜、气雾剂制剂和玻璃圆筒的频繁污染证明污染是由耦合液引起的。还发现超声波对气雾剂制剂中的微生物有破坏作用。调查表明,每次使用前都应更换吸入面罩及其与设备相连的管道,而雾化器的其他部件和气雾剂制剂每吸入15次可更换一次。还指出,用2%的戊二醛水溶液(室温下30分钟)对设备的不同部件进行消毒,效率约为95%。