Szivek J A, Magee F P
Harrington Arthritis Research Center, Phoenix, Arizona.
J Invest Surg. 1989;2(2):195-206. doi: 10.3109/08941938909015351.
Bone remodeling resulting from implant insertion has been attributed to changes in the bone's strain state. Since remodeling takes several months, it was this study's purpose to develop a long-term in vivo strain sensor. Porous surfaced metal tabs were attached to a standard strain gauge. Two standard gauges and the porous tabbed gauge were attached to one femur and three standard gauges to the contralateral femur of a greyhound. Tissue ingrowth provided an attachment mechanism for the porous tabbed gauge in vivo. Gauge measurements were compared to those from the standard gauges. Post sacrifice testing allowed further comparisons. After histological preparation the femoral section shapes as well as the gauge locations were examined and photographed. The porous tabbed gauge remained bonded and sensed strain throughout the 8-week implantation period, while the standard gauges debonded and were unable to detect strain after 3 weeks. During testing, the measurements from the porous tabbed gauge were lower than those from the standard gauges. This was consistent with the histology which indicated that fibrous tissue had invaded the gauge's porous surface. Although the use of tissue ingrowth as an attachment mechanism seems to be worthwhile, a means of insuring bone ingrowth is necessary.
植入物插入导致的骨重塑归因于骨应变状态的变化。由于重塑需要数月时间,本研究的目的是开发一种长期的体内应变传感器。多孔表面金属片附着在标准应变片上。将两个标准应变片和多孔带片应变片附着在一只灵缇犬的一根股骨上,将三个标准应变片附着在对侧股骨上。组织向内生长为多孔带片应变片在体内提供了附着机制。将应变片的测量结果与标准应变片的测量结果进行比较。处死动物后的测试允许进行进一步比较。在组织学制备后,检查并拍摄股骨切片形状以及应变片位置。在整个8周的植入期内,多孔带片应变片保持粘结并能感知应变,而标准应变片在3周后脱粘且无法检测到应变。在测试期间,多孔带片应变片的测量值低于标准应变片的测量值。这与组织学结果一致,组织学结果表明纤维组织已侵入应变片的多孔表面。尽管利用组织向内生长作为附着机制似乎是值得的,但确保骨向内生长的方法是必要的。