[运用“T-Scan 系统”进行咬合分析的研究。1. 其测量准确性]

[Study on occlusal analysis by means of 'T-Scan system'. 1. Its accuracy for measurement].

作者信息

Tokumura K, Yamashita A

出版信息

Nihon Hotetsu Shika Gakkai Zasshi. 1989 Oct;33(5):1037-43. doi: 10.2186/jjps.33.1037.

Abstract

The T-Scan system which was developed by Dr. W.L. Maness in 1987 is a new computerized device which can analyze time and force of occlusal contacts. It has potential to analyze occlusal contacts quantitatively. The purpose of this study is to investigate its accuracy for measurement as quantitative analysis device. The following conclusions can be drawn from this study: 1. The time display of T-Scan system was approximately 64/100 of real time. 2. The force display of T-Scan system showed weak correlation to loads when loads from 1 to 10 kg were added on the sensor at random position. 3. T-Scan sensor had non-sensible areas caused by its structure.

摘要

由W.L. 马内斯博士于1987年研发的T-Scan系统是一种新型的计算机化设备,它能够分析咬合接触的时间和力量。它有对咬合接触进行定量分析的潜力。本研究的目的是调查其作为定量分析设备在测量方面的准确性。从本研究中可以得出以下结论:1. T-Scan系统的时间显示约为实际时间的64/100。2. 当在传感器的随机位置施加1至10千克的负荷时,T-Scan系统的力量显示与负荷的相关性较弱。3. T-Scan传感器因其结构存在无感应区域。