两种用于测量绷带下压力的设备的实际局限性:对临床实践的启示。

Practical limitations of two devices used for the measurement of sub-bandage pressure: implications for clinical practice.

作者信息

Thomas S

机构信息

BPharm, PhD, Medical Device Consultancy, MedetecTM.

出版信息

J Wound Care. 2014 Jun;23(6):300, 302-6, 308-9 passim. doi: 10.12968/jowc.2014.23.6.300.

Abstract

OBJECTIVE

This study was undertaken to examine the accuracy of two hand-held devices commonly used to measure the pressures produced by extensible bandages.

METHOD

The performance of the pneumatic sensors of two devices, the Kikuhime and Picopress instruments was first examined in air in a compression chamber, then subsequently beneath multiple layers of bandages applied to standard cylinders with predetermined levels of tension.

RESULTS

In the compression chamber, both instruments provided readings that were typically within 1 mmHg of the reference value, but on curved formers in free air or beneath bandages the accuracy of both sensors was greatly reduced, influenced both by the curvature of the cylinders and the volume of air contained in the sensor capsule. The Picopress instrument recorded pressures up to 70% higher than predicted, particularly on the smaller cylinders and at the lower end of the pressure range (circa 20mmHg). At 40mmHg measured pressure were around 40% higher than predicted values in some instances. The accuracy of the Kikuhime was greatly influenced by the calibration technique but percentage deviations as high as 150% were recorded in some tests.

CONCLUSION

Pneumatic pressure sensors used to record the pressures developed beneath compression bandages are much less accurate than is commonly believed. Calibration studies using air chambers or water tanks have no relevance to the normal clinical use of these devices. These limitations should be considered by clinicians when making judgments about the performance of other medical devices such as bandages or stockings. They also call into question the validity of many published studies which rely upon such measurements.

DECLARATION OF INTEREST

This project was generously supported by 3M Healthcare who provided a small research grant to assist with the costs associated with the construction of the test equipment.

摘要

目的

本研究旨在检验两种常用于测量可伸展绷带产生压力的手持设备的准确性。

方法

首先在压缩室内的空气中检测两种设备(菊姬仪器和皮可压力仪器)的气动传感器性能,随后在施加了具有预定张力水平的多层绷带的标准圆柱体下方进行检测。

结果

在压缩室内,两种仪器提供的读数通常与参考值相差在1 mmHg以内,但在自由空气中或绷带下方的弯曲成型器上,两种传感器的准确性都大大降低,这受到圆柱体曲率和传感器胶囊内空气体积的影响。皮可压力仪器记录的压力比预测值高出70%,特别是在较小的圆柱体上以及压力范围的低端(约20 mmHg)。在某些情况下,测量压力为40 mmHg时比预测值高出约40%。菊姬仪器的准确性受校准技术的影响很大,但在一些测试中记录到的偏差百分比高达150%。

结论

用于记录压缩绷带下方产生压力的气动压力传感器的准确性远低于普遍认为的水平。使用气室或水箱进行的校准研究与这些设备的正常临床使用无关。临床医生在判断绷带或长袜等其他医疗设备的性能时应考虑这些局限性。它们还对许多依赖此类测量的已发表研究的有效性提出了质疑。

利益声明

本项目得到了3M医疗保健公司的慷慨支持,该公司提供了一笔小额研究资助,以协助支付与测试设备构建相关的费用。