Saito T, Chaplin D J, Lam G K, Skarsgard L D
Medical Biophysics Unit, British Columbia Cancer Research Centre, Vancouver, Canada.
Radiat Res. 1989 Jan;117(1):8-20.
The survival of mouse jejunal crypt cells was measured using a microcolony assay following irradiation with one of three different pion stopping volumes. All treatments were done using the beam spot scanning system developed for clinical therapy at TRIUMF. Treatment volumes were Small, 10 ml of volume using 4 beam spots with a narrow stopping peak; Moderate, 40 ml of volume using 4 beam spots with an extended peak; Large, 1.2 liters of volume using 37 spots with an extended peak. The dose rate fluctuated considerably during treatment because of the scanning procedure. The relative biological effectiveness (RBE) values of pions were 1.11, 1.04, and 1.16 for the small, moderate, and large volumes, respectively. As the width of the stopping peak increased, from the narrow 4-cm peak of the small volume to the 12-cm peak of the moderate volume, the effectiveness of the pion beam decreased. This presumably reflects the low linear energy transfer contribution of the large number of passing pions in the wider stopping peak. The results indicate, however, a greater biological effectiveness for the large volume than for the moderate volume. Even though these fields had stopping peaks of similar width, the field size and dose rate were different. Although the increase in RBE may be partly due to an increased neutron flux from the larger volume, our results suggest that much of this apparent increase could be explained by differences in dose rate as a function of time for pions versus X rays.
在用三种不同的π介子阻止体积之一进行辐照后,使用微集落测定法测量小鼠空肠隐窝细胞的存活率。所有治疗均使用为TRIUMF临床治疗开发的束斑扫描系统进行。治疗体积为:小体积,使用4个束斑,体积为10 ml,具有窄的阻止峰;中等体积,使用4个束斑,体积为40 ml,具有扩展峰;大体积,使用37个束斑,体积为1.2升,具有扩展峰。由于扫描过程,治疗期间剂量率波动很大。对于小体积、中等体积和大体积,π介子的相对生物效应(RBE)值分别为1.11、1.04和1.16。随着阻止峰宽度的增加,从小体积的窄4厘米峰到中等体积的12厘米峰,π介子束的有效性降低。这可能反映了较宽阻止峰中大量通过的π介子的低传能线密度贡献。然而,结果表明大体积的生物有效性高于中等体积。尽管这些照射野的阻止峰宽度相似,但照射野大小和剂量率不同。虽然RBE的增加可能部分归因于较大体积产生的中子通量增加,但我们的结果表明,这种明显增加的很大一部分可以用π介子与X射线的剂量率随时间的差异来解释。