Studzińska Sylwia, Buszewski Bogusław
Faculty of Chemistry, Nicolaus Copernicus University, 7 Gagarin Str., 87-100, Toruń, Poland,
Anal Bioanal Chem. 2014 Nov;406(28):7127-36. doi: 10.1007/s00216-014-7959-5. Epub 2014 Jun 18.
Ultra high-performance [corrected] liquid chromatography has been used for the separation and analysis of unmodified and modified antisense oligonucleotides. For this reason, we tested various columns of low particle sizes in our analysis of unmodified and phosphorothioate oligonucleotides. The influence of both the type and concentration of ion-pair reagent on the retention of the studied biomolecules was tested. The developed methods were used for separation of unmodified oligonucleotides and to determine antisense oligonucleotides in human serum samples. The results proved that octadecyl and phenyl columns are the most selective in the resolution of oligonucleotides which differ in the position of single nucleotides in the sequence. The phenyl column was selected and applied for the analysis of phosphorothioate oligonucleotides in serum samples. The calibration plots showed good linearity within the test concentration ranges. The intra-day CV of the calibration curve slopes was in the range of 1.6 to 4.2 %. The limits of detection (LODs) were in the range of 0.11-0.16 μg mL(-1), while the limit of quantification (LOQ) values were between 0.35 and 0.51 μg mL(-1).
超高效液相色谱已用于未修饰和修饰的反义寡核苷酸的分离与分析。因此,在对未修饰和硫代磷酸酯寡核苷酸的分析中,我们测试了各种低粒径的色谱柱。测试了离子对试剂的类型和浓度对所研究生物分子保留的影响。所开发的方法用于分离未修饰的寡核苷酸并测定人血清样品中的反义寡核苷酸。结果证明,十八烷基柱和苯基柱在分离序列中单个核苷酸位置不同的寡核苷酸时具有最高的选择性。选择苯基柱用于血清样品中硫代磷酸酯寡核苷酸的分析。校准曲线在测试浓度范围内显示出良好的线性。校准曲线斜率的日内变异系数在1.6%至4.2%范围内。检测限(LOD)在0.11 - 0.16μg mL(-1)范围内,而定量限(LOQ)值在0.35至0.51μg mL(-1)之间。