Liu Yawei, Tian Miao, Jiang Xinzhi, Luo Jiakun, Yi Zhiyong
Health Division, 3305 Hospital, Guard Bureau of General Staff Department of PLA, Beijing 100017, China.E-mail:
Nan Fang Yi Ke Da Xue Xue Bao. 2014 Jun;34(6):901-3.
To evaluation the performance of a heart-type fatty acid-binding protein (H-FABP) ELISA detection kit in the clinical diagnosis of acute coronary syndrome (ACS).
Plasma or serum samples from 160 suspected ACS patients hospitalized in General Hospital of PLA were examined using Lanzhou H-FABP reagent kit and Holland H-FABP kit. Correlation of the two kits was evaluated and Kappa test was used to examine the consistency of the results of the two products.
The sensitivity of H-FABP diagnosis of ACS detection Lanzhou kit was 91.8%, the specificity was 88.7%, and the total diagnostic rate was 90.42%. The sensitivity of H-FABP diagnosis of ACS detection Holland kit was 90.3%, the specificity was 86.8%, and the total diagnostic rate was 88.75%. The test results showed that two products yielded comparable results (P=0.668, >0.05) with a good consistency (Kappa=0.726, P<0.01).
H-FABP ELISA detection kit produccted by LanZhou biological research institute has a good correlation with H-FABP detection kit produced by HBT company of Holland and has the potential for clinical application.