Talmo Carl T, Sharp Kinzie G, Malinowska Magdalena, Bono James V, Ward Daniel M, LaReau Justin
Orthopedics. 2014 Jun;37(6):e592-5. doi: 10.3928/01477447-20140528-62.
Modular femoral heads have been used successfully for many years in total hip arthroplasty. Few complications have been reported for the modular Morse taper connection between the femoral head and trunnion of the stem in metal-on-polyethylene bearings. Although there has always been some concern over the potential for fretting, corrosion, and generation of particulate debris at the modular junction, this was not considered a significant clinical problem. More recently, concern has increased because fretting and corrosive debris have resulted in rare cases of pain, adverse local tissue reaction, pseudotumor, and osteolysis. Larger femoral heads, which have gained popularity in total hip arthroplasty, are suspected to increase the potential for local and systemic complications of fretting, corrosion, and generation of metal ions because of greater torque at the modular junction. A less common complication is dissociation of the modular femoral heads. Morse taper dissociation has been reported in the literature, mainly in association with a traumatic event, such as closed reduction of a dislocation or fatigue fracture of the femoral neck of a prosthesis. This report describes 3 cases of spontaneous dissociation of the modular prosthetic femoral head from the trunnion of the same tapered titanium stem because of fretting and wear of the Morse taper in a metal-on-polyethylene bearing. Continued clinical and scientific research on Morse taper junctions is warranted to identify and prioritize implant and surgical factors that lead to this and other types of trunnion failure to minimize complications associated with Morse taper junctions as hip implants and surgical techniques continue to evolve.
模块化股骨头已成功应用于全髋关节置换术多年。在金属对聚乙烯关节中,关于股骨头与柄的耳轴之间的模块化莫氏锥度连接,所报告的并发症很少。尽管人们一直担心在模块化连接处可能出现微动、腐蚀以及产生颗粒碎片,但这在临床上并不被视为重大问题。最近,这种担忧有所增加,因为微动和腐蚀性碎片已导致罕见的疼痛、局部组织不良反应、假肿瘤和骨溶解病例。在全髋关节置换术中越来越受欢迎的较大尺寸股骨头,由于在模块化连接处产生更大的扭矩,被怀疑会增加微动、腐蚀和金属离子产生的局部及全身并发症的可能性。一种不太常见的并发症是模块化股骨头的分离。文献中已报道过莫氏锥度分离,主要与创伤性事件有关,比如脱位的闭合复位或假体股骨颈的疲劳骨折。本报告描述了3例在金属对聚乙烯关节中,由于莫氏锥度的微动和磨损,模块化人工股骨头从同一锥形钛柄的耳轴上自发分离的病例。随着髋关节植入物和手术技术不断发展,有必要继续对莫氏锥度连接进行临床和科学研究,以确定导致这种及其他类型耳轴失效的植入物和手术因素,并对其进行优先级排序,从而将与莫氏锥度连接相关的并发症降至最低。