Machado Priscilla, Eisenbrey John R, Cavanaugh Barbara, Forsberg Flemming
Department of Radiology, Thomas Jefferson University, Philadelphia, Pennsylvania, USA.
Department of Radiology, Thomas Jefferson University, Philadelphia, Pennsylvania, USA.
Ultrasound Med Biol. 2014 Sep;40(9):2321-4. doi: 10.1016/j.ultrasmedbio.2014.04.008. Epub 2014 Jul 9.
A new commercial image processing technique (MicroPure, Toshiba America Medical Systems, Tustin, CA, USA) that identifies breast microcalcifications was evaluated at the time of patients' annual screening mammograms. Twenty women scheduled for annual screening mammography were enrolled in the study. Patients underwent bilateral outer-upper-quadrant real-time dual gray scale ultrasound and MicroPure imaging using an Aplio XG scanner (Toshiba). MicroPure combines non-linear imaging and speckle suppression to mark suspected calcifications as white spots in a blue overlay image. Four independent and blinded readers analyzed digital clips to determine the presence or absence of microcalcifications and artifacts. The presence of microcalcifications determined by readers was not significantly different from that of mammography (p = 0.57). However, the accuracy was low overall (52%) and also in younger women (<50 years, 54%). In conclusion, although microcalcifications can be identified using MicroPure imaging, this method is not currently appropriate for a screening population and should be used in more focused applications.
一种用于识别乳腺微钙化的新型商业图像处理技术(美国加利福尼亚州图斯廷市东芝美国医疗系统公司的MicroPure)在患者年度乳腺钼靶筛查时进行了评估。二十名计划进行年度乳腺钼靶筛查的女性被纳入该研究。患者接受了双侧外上象限实时双灰阶超声检查,并使用Aplio XG扫描仪(东芝)进行了MicroPure成像。MicroPure结合了非线性成像和斑点抑制技术,在蓝色叠加图像中将疑似钙化标记为白点。四名独立且不知情的阅片者分析数字片段,以确定微钙化和伪影的存在与否。阅片者确定的微钙化存在情况与乳腺钼靶检查的结果无显著差异(p = 0.57)。然而,总体准确性较低(52%),在年轻女性(<50岁,54%)中也是如此。总之,虽然使用MicroPure成像可以识别微钙化,但该方法目前不适用于筛查人群,应在更有针对性的应用中使用。