Peake Roy W A, Hartigan Christina R, Esposito Christopher L, Kellogg Mark D, Gabler Jessica, Wang Sihe, Breaud Autumn, Di Bussolo Joseph, Moskowitz Joshua, Clarke William
*Department of Laboratory Medicine, Boston Children's Hospital, MA; †Thermo Scientific, Franklin, MA; ‡Department of Clinical Pathology, Cleveland Clinic LL Lab, OH; and §Department of Pathology, Johns Hopkins University Hospital, Baltimore, MD.
Ther Drug Monit. 2015 Apr;37(2):161-71. doi: 10.1097/FTD.0000000000000120.
Immunosuppressant drugs (ISDs) are commonly prescribed to solid organ transplant patients. Their narrow therapeutic index and potential for toxicity necessitates careful monitoring of blood concentrations. Liquid chromatography-tandem mass spectrometry (LC-MS/MS) methods are increasingly used for ISD measurement. However, there remain many challenges with this methodology, particularly regarding interassay variability. The Thermo Scientific Prelude is an online extraction/liquid chromatography platform that uses turbulent flow technology coupled with MS/MS. A multicenter evaluation of the Prelude for the measurement of cyclosporine A, tacrolimus, and sirolimus is described. ISDs were measured at each site using standardized protocols. Sample preparation liquid chromatography-MS/MS was performed using the Prelude coupled to a TSQ Vantage. Chromatography was achieved with a Cyclone-P TurboFlow/Accucore C8 column combination using a multisolvent loading and eluting pump system. Mass spectrometry acquisitions were performed in selective reaction monitoring mode and data processed using TraceFinder (version 3.1). Multisite mean imprecision for cyclosporine A ranged from 8.8% (54 mcg/L) to 9.8% (450 mcg/L); for tacrolimus, 4.7% (15.5 mcg/L) to 12.6% (2.5 mcg/L); for sirolimus, 7.4% (19.9 mcg/L) to 16.5% (2.6 mcg/L). Approximately 110 specimens were used for method comparison. For cyclosporine A, mean bias against the multisite mean ranged from -18% to 1%; for tacrolimus, values ranged from -7% to 4%; for sirolimus, values ranged from -4% to 2%. Comparisons of multisite mean Prelude results with routine ISD method results was also performed for cyclosporine A (slope = 0.7878, intercept = 24.16, r = 0.98), tacrolimus (slope = 0.9391, intercept = 0.1017, r = 98), and sirolimus (slope = 0.9618, intercept = 0.1483, r = 0.97). The Prelude ISD method offers acceptable and comparable multisite performance. This study has also highlighted the importance of adopting standardized protocols and LC-MS/MS methods for better comparability between ISD assays.
免疫抑制药物(ISDs)通常会开给实体器官移植患者。其治疗指数狭窄且具有潜在毒性,因此需要仔细监测血药浓度。液相色谱 - 串联质谱(LC - MS/MS)方法越来越多地用于ISD测量。然而,这种方法仍存在许多挑战,特别是在批间变异性方面。赛默飞世尔科技的Prelude是一个在线萃取/液相色谱平台,它使用湍流技术与MS/MS联用。本文描述了对Prelude用于测量环孢素A、他克莫司和西罗莫司的多中心评估。在每个站点使用标准化方案测量ISDs。使用与TSQ Vantage联用的Prelude进行样品制备液相色谱 - MS/MS分析。使用Cyclone - P TurboFlow/Accucore C8柱组合,通过多溶剂加载和洗脱泵系统实现色谱分离。质谱采集在选择性反应监测模式下进行,数据使用TraceFinder(版本3.1)处理。环孢素A的多站点平均不精密度范围为8.8%(54 mcg/L)至9.8%(450 mcg/L);他克莫司为4.7%(15.5 mcg/L)至12.6%(2.5 mcg/L);西罗莫司为7.4%(19.9 mcg/L)至16.5%(2.6 mcg/L)。大约110个标本用于方法比较。对于环孢素A,相对于多站点平均值的平均偏差范围为 - 18%至1%;对于他克莫司,偏差值范围为 - 7%至4%;对于西罗莫司,偏差值范围为 - 4%至2%。还对环孢素A(斜率 = 0.7878,截距 = 24.16,r = 0.98)、他克莫司(斜率 = 0.9391,截距 = 0.1017,r = 0.98)和西罗莫司(斜率 = 0.9618,截距 = 0.1483,r = 0.97)进行了Prelude多站点平均结果与常规ISD方法结果的比较。Prelude ISD方法提供了可接受的且具有可比性的多站点性能。本研究还强调了采用标准化方案和LC - MS/MS方法对于提高ISD检测之间可比性的重要性。