Eskandarlou Amir, Jafari Amir Abbas, Mohammadi Mohammad, Zehtabian Mehdi, Faghihi Reza, Shokri Abbas, Pourolajal Jalal
Department of Radiology, Faculty of Dentistry, Hamadan University of Medical sciences, Hamadan, Iran.
Department of Medical Physics, Faculty of Medicine, Hamadan University of Medical sciences, Hamadan, Iran Department of Medical Physics, Royal Adelaide Hospital, Adelaide, South Australia, Australia.
J Xray Sci Technol. 2014;22(4):459-70. doi: 10.3233/XST-140438.
The effectiveness of added filters including conventional and rare earth materials for dental radiography tasks was investigated using a simulation approach. Current study focuses on the combination of a range of various filters to investigate the reduction of radiation absorbed dose and improving the quality of a radiography image. To simulate the X-ray beam spectrum, a MCNP5 code was applied. Relative intensity, beam quality, and mean energy were investigated for a typical dental radiography machine. The impact of different rare-earth materials with different thicknesses and tube voltages on the X-ray spectrum was investigated. For Aluminum as a conventional filter, the modeled X-ray spectra and HVL values were in a good agreement with those reported by IPEM. The results showed that for a 70 kVp voltage, with an increase of the thickness and atomic number of a given added filters, an increase of HVL values were observed. However, with the increase of the attenuator thickness, X-ray beam intensity decreases. For mean energy, different results were observed. It was also found that rare earth made filters reduce high energy X-ray radiation due to k-edge absorption. This leads to an ideal beam for intra-oral radiography tasks. However, as a disadvantage of rare earth added filters, the reduction of the tube output levels should also be considered.
采用模拟方法研究了包括传统材料和稀土材料在内的附加滤过器在牙科放射摄影任务中的有效性。当前研究聚焦于一系列不同滤过器的组合,以研究辐射吸收剂量的降低以及改善放射摄影图像的质量。为了模拟X射线束光谱,应用了MCNP5代码。对一台典型的牙科放射摄影设备的相对强度、射线质和平均能量进行了研究。研究了不同厚度和管电压的不同稀土材料对X射线光谱的影响。对于作为传统滤过器的铝,模拟的X射线光谱和半值层(HVL)值与IPEM报告的结果吻合良好。结果表明,对于70 kVp的电压,随着给定附加滤过器厚度和原子序数的增加,观察到HVL值增加。然而,随着衰减器厚度的增加,X射线束强度降低。对于平均能量,观察到了不同的结果。还发现稀土制成的滤过器由于K边吸收而减少了高能X射线辐射。这产生了用于口腔内放射摄影任务的理想射线束。然而,作为添加稀土滤过器的一个缺点,管输出水平的降低也应予以考虑。