Tyndall D A
Oral Surg Oral Med Oral Pathol. 1986 Aug;62(2):205-11. doi: 10.1016/0030-4220(86)90047-2.
A laboratory investigation was carried out to assess the effect of various types of rare earth filter materials on the energy spectrum and concomitant reduced exposure values of diagnostic x-ray beams at 70, 80, and 90 kVp. An x-ray spectroscope was constructed and used to generate the energy spectra of beams passing through the various rare earth filter materials. Photographs were made of each spectrum, and live-time gross photon counts were recorded. Following spectral determinations, ionization chamber readings were generated for each filter material. Substantial effects on x-ray spectra and reduction of exposure values were noted. The degree of these effects were dependent on the atomic number, k-edge, and thickness of each filter. Metallic forms of rare earth materials proved to be more effective than the salt forms with erbium offering the greatest potential for reduction in exposures over the range of experimental kilovolt (peak) values.
进行了一项实验室研究,以评估各种类型的稀土过滤材料对70、80和90千伏峰值下诊断性X射线束的能谱及相应降低的照射量值的影响。构建了一台X射线光谱仪,并用于生成穿过各种稀土过滤材料的射线束的能谱。对每个能谱进行拍照,并记录实时总光子计数。在进行光谱测定之后,针对每种过滤材料生成电离室读数。注意到对X射线光谱和照射量值降低有显著影响。这些影响的程度取决于每种过滤材料的原子序数、K吸收边和厚度。稀土材料的金属形式被证明比盐形式更有效,在实验千伏(峰值)值范围内,铒在降低照射量方面具有最大潜力。