Battmer Rolf-Dieter, Borel Stephanie, Brendel Martina, Buchner Andreas, Cooper Huw, Fielden Claire, Gazibegovic Dzemal, Goetze Romy, Govaerts Paul, Kelleher Katherine, Lenartz Thomas, Mosnier Isabelle, Muff Joanne, Nunn Terry, Vaerenberg Bart, Vanat Zebunissa
Cochlear Implants Int. 2015 Mar;16(2):100-9. doi: 10.1179/1754762814Y.0000000093. Epub 2014 Aug 12.
To compare the fitting time requirements and the efficiency in achieving improvements in speech perception during the first 6 months after initial stimulation of computer-assisted fitting with the Fitting to Outcome eXpert' (FOX) and a standard clinical fitting procedure.
Twenty-seven post-lingually deafened adults, newly implanted recipients of the Advanced Bionics HiRes 90K™ cochlear implant from Germany, the UK, and France took part in a controlled, randomized, clinical study. Speech perception was measured for all participants and fitting times were compared across groups programmed using FOX and conventional programming methods.
The fitting time for FOX was significantly reduced at 14 days (P < 0.001) but equivalent over the 6-month period. The groups were not well matched for duration of deafness; therefore, speech perception could not be compared across groups.
Despite including more objective measures of performance than a standard fitting approach and the adjustment of a greater range of parameters during initial fitting, FOX did not add to the overall fitting time when compared to the conventional approach. FOX significantly reduced the fitting time in the first 2 weeks and by providing a standard fitting protocol, reduced variability across centres.
FOX computer-assisted fitting can be successfully used at switch on, in different clinical environments, reducing fitting time in the first 2 weeks and is efficient at providing a usable program.
比较使用“适配至结果专家”(FOX)计算机辅助适配与标准临床适配程序在初次刺激后的前6个月内,实现言语感知改善所需的适配时间及效率。
27名语后聋成年人,他们是来自德国、英国和法国的Advanced Bionics HiRes 90K™人工耳蜗新植入者,参与了一项对照、随机临床研究。对所有参与者进行言语感知测量,并比较使用FOX和传统编程方法进行编程的各组的适配时间。
FOX在14天时的适配时间显著缩短(P < 0.001),但在6个月期间两者相当。各组在耳聋持续时间上匹配不佳;因此,无法在组间比较言语感知情况。
尽管与标准适配方法相比,FOX纳入了更多客观的性能指标,且在初次适配期间调整了更大范围的参数,但与传统方法相比,FOX并未增加总体适配时间。FOX显著缩短了前2周的适配时间,并通过提供标准适配方案,减少了各中心之间的差异。
FOX计算机辅助适配可在开机时成功应用于不同临床环境,减少前2周的适配时间,并且能有效地提供可用程序。