Programa de Pós-Graduação em Geociências e Meio Ambiente, Instituto de Geociências e Ciências Exatas, UNESP - Univ. Estadual Paulista, Av. 24-A, 1515, CEP 13506-900, Rio Claro, SP, Brazil; Centro de Estudos Ambientais, UNESP - Univ. Estadual Paulista, Av. 24-A, 1515, CEP 13506-900, Rio Claro, SP, Brazil.
Centro de Estudos Ambientais, UNESP - Univ. Estadual Paulista, Av. 24-A, 1515, CEP 13506-900, Rio Claro, SP, Brazil.
Talanta. 2014 Nov;129:417-21. doi: 10.1016/j.talanta.2014.05.025. Epub 2014 May 24.
In this work, a device based on diffusive gradients in thin films (DGT) was evaluated for the determination of Hg(II) in river water. The DGT device was assembled with a cellulose phosphate ion exchange membrane (P81 Whatman) as a binding phase and agarose gel 1.5% (m/v) as a diffusive layer. Laboratory deployments showed that the binding of Hg(2+) ([Hg(DGT)]/[Hg(solution)]) by P81 membrane was more effective (97%) than the Chelex 100 resin (80%).The effect of ionic strength, pH and potential interfering ions on Hg binding with DGT׳s was investigated. The results showed no significant effect on the binding of Hg(II) at pH range from 3.5 to 8.5 and at an ionic strength range from 0.0005 to 0.1 mol L(-1). Uptakes of 50 µg L(-1) Hg(II) by P81 membrane were not affected by Fe, Mn, Zn, Cu, Ca and Mg at the concentration range of 200-1800 µg L(-1). Finally, the DGT device using the P81 as the binding layer was applied for in situ measurements of Hg in river water. For in situ measurements, the labile Hg concentration (from <2 to 13 ng L(-1)) was lower than 10% of the dissolved fraction (from 155 to 446 ng L(-1)).
在本研究中,对一种基于薄膜扩散梯度(DGT)的装置进行了评估,用于测定河水中的汞(II)。DGT装置由磷酸纤维素离子交换膜(Whatman P81)作为结合相和1.5%(m/v)的琼脂糖凝胶作为扩散层组装而成。实验室部署表明,P81膜对汞(2+)的结合([Hg(DGT)]/[Hg(溶液)])比Chelex 100树脂(80%)更有效(97%)。研究了离子强度、pH值和潜在干扰离子对汞与DGT结合的影响。结果表明,在pH值范围为3.5至8.5以及离子强度范围为0.0005至0.1 mol L(-1)时,对汞(II)的结合没有显著影响。在200-1800 µg L(-1)的浓度范围内,P81膜对50 µg L(-1)汞(II)的摄取不受铁、锰、锌、铜、钙和镁的影响。最后,将使用P81作为结合层的DGT装置应用于河水中汞的原位测量。对于原位测量,不稳定汞浓度(<2至13 ng L(-1))低于溶解部分(155至446 ng L(-1))的10%。