Tafurt-Cardona Makenly, Eismann Carlos Eduardo, Suárez Carlos Alfredo, Menegário Amauri Antonio, Silva Luko Karen, Sargentini Junior Ézio
Programa de Pós-graduação em Geociências e Meio Ambiente, Instituto de Geociências e Ciências Exatas, UNESP - Univ. Estadual Paulista, Av. 24-A, 1515, CEP: 13506-900, Rio Claro, SP, Brazil; Centro de Estudos Ambientais, UNESP - Univ. Estadual Paulista, Av. 24-A, 1515, CEP: 13506-900, Rio Claro, SP, Brazil.
Centro de Estudos Ambientais, UNESP - Univ. Estadual Paulista, Av. 24-A, 1515, CEP: 13506-900, Rio Claro, SP, Brazil.
Anal Chim Acta. 2015 Aug 5;887:38-44. doi: 10.1016/j.aca.2015.07.035. Epub 2015 Aug 10.
Saccharomyces cerevisiae immobilized in agarose gel as binding phase and polyacrylamide as diffusive layer in the diffusive gradient in thin films technique (DGT) was used for selective determination of methylmercury (MeHg). Deployment tests showed good linearity in mass uptake up to 48 h (3276 ng). When coupling the DGT technique with Cold Vapor Atomic Fluorescence Spectrometry, the method has a limit of detection of 0.44 ng L(-1) (pre concentration factor of 11 for 48 h deployment). Diffusion coefficient of 7.03 ± 0.77 × 10(-6) cm(2) s(-1) at 23 °C in polyacrylamide gel (pH = 5.5 and ionic strength = 0.05 mol L(-1) NaCl) was obtained. Influence of ionic strength (from 0.0005 mol L(-1) to 0.1 mol L(-1) NaCl) and pH (from 3.5 to 8.5) on MeHg uptake were evaluated. For these range, recoveries of 84-105% and 84-98% were obtained for ionic strength and pH respectively. Potential interference due to presence of Cu, Fe, Mn, Zn was also assessed showing good recoveries (70-87%). The selectivity of the proposed approach was tested by deployments in solutions containing MeHg and Hg(II). Results obtained showed recoveries of 102-115 % for MeHg, while the uptake of Hg(II) was insignificant. The proposed approach was successfully employed for in situ measurements in the Negro River (Manaus-AM, Brazil).
将固定在琼脂糖凝胶中的酿酒酵母作为结合相,聚丙烯酰胺作为薄膜扩散梯度技术(DGT)中的扩散层,用于选择性测定甲基汞(MeHg)。部署测试表明,在长达48小时(3276纳克)的质量摄取中具有良好的线性关系。当将DGT技术与冷蒸气原子荧光光谱法结合使用时,该方法的检测限为0.44纳克/升(48小时部署的预浓缩因子为11)。在23℃下,聚丙烯酰胺凝胶(pH = 5.5,离子强度 = 0.05摩尔/升NaCl)中的扩散系数为7.03±0.77×10⁻⁶平方厘米/秒。评估了离子强度(从0.0005摩尔/升到0.1摩尔/升NaCl)和pH值(从3.5到8.5)对MeHg摄取的影响。在此范围内,离子强度和pH值的回收率分别为84 - 105%和84 - 98%。还评估了铜、铁、锰、锌存在时的潜在干扰,回收率良好(70 - 87%)。通过在含有MeHg和Hg(II)的溶液中进行部署,测试了所提出方法的选择性。所得结果表明,MeHg的回收率为102 - 115%,而Hg(II)的摄取量微不足道。所提出的方法已成功用于巴西马瑙斯-亚马逊州内格罗河的原位测量。