术中 HyperEye Medical System 血管造影术在冠状动脉旁路移植术中的疗效

Efficacy of intraoperative HyperEye Medical System angiography for coronary artery bypass grafting.

作者信息

Yamamoto Masaki, Orihashi Kazumasa, Nishimori Hideaki, Handa Takemi, Kondo Nobuo, Fukutomi Takashi, Sato Takayuki

机构信息

Department of Surgery 2, Kochi Medical School, Kohasu, Oko-cho, Nankoku, Kochi, 783-8505, Japan,

出版信息

Surg Today. 2015 Aug;45(8):966-72. doi: 10.1007/s00595-014-1015-0. Epub 2014 Aug 28.

Abstract

PURPOSE

When performing coronary artery bypass grafting, anastomotic insufficiency needs to be detected during surgery. We developed a novel indocyanine green angiography system, the HyperEye Medical System (HEMS), which enables color imaging of the bypass flow. This article described the accuracy of HEMS angiography for predicting graft patency.

METHODS

A total of 144 grafts in 40 coronary artery bypass grafting patients were assessed by HEMS angiography, a transit time flowmeter (TTF) during surgery and fluoroscopic angiography 1 year after the operation.

RESULTS

HEMS angiography showed normal flow in 133 grafts, but abnormal flow in 11. Fluoroscopic angiography showed that 130 of the 133 "normal" grafts were patent (negative predictive value: 97.7%) and that nine of the 11 "abnormal" grafts were occluded (positive predictive value: 81.8%). As a result, 134 grafts were assessed as normal and 10 as abnormal by TTF. Fluoroscopic angiography showed that 124 of these 134 grafts were patent (negative predictive value: 92.5%), whereas two of the 10 grafts were occluded (positive predictive value: 20.0%). For the 127 grafts with compatible results by the HEMS and TTF assessments, the positive and negative predictive values were 100 and 97.6%, respectively.

CONCLUSION

HEMS angiography of a bypass graft may provide an accurate prediction of the graft patency after surgery.

摘要

目的

在进行冠状动脉旁路移植术时,手术过程中需要检测吻合口是否存在功能不全。我们开发了一种新型吲哚菁绿血管造影系统——HyperEye Medical System(HEMS),它能够对搭桥血流进行彩色成像。本文描述了HEMS血管造影预测移植血管通畅性的准确性。

方法

对40例冠状动脉旁路移植术患者的144条移植血管进行了评估,评估方法包括手术期间的HEMS血管造影、通过渡时间血流仪(TTF)以及术后1年的荧光透视血管造影。

结果

HEMS血管造影显示133条移植血管血流正常,11条异常。荧光透视血管造影显示,133条“正常”移植血管中有130条通畅(阴性预测值:97.7%),11条“异常 ”移植血管中有9条闭塞(阳性预测值:81.8%)。结果,通过TTF评估,134条移植血管为正常,10条为异常。荧光透视血管造影显示,这134条移植血管中有124条通畅(阴性预测值:92.5%),而10条移植血管中有2条闭塞(阳性预测值:20.0%)。对于HEMS和TTF评估结果一致的127条移植血管,其阳性和阴性预测值分别为100%和97.6%。

结论

对移植血管进行HEMS血管造影可以准确预测术后移植血管的通畅情况。